Skip to content

Indeflator recallZ-0925-2022 · Abbott Vascular · initiated Mar 11, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Vascular or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to an increase in complaint trend for leaks and intermittent/loose connections.

FDA root cause category: Component change control

What the company is doing

On 03/11/2022, the firm sent an "URGENT FIELD SAFETY NOTICE/DEVICE RECALL" Letter via visit, email or mail, to customers informing them that specific lots of 20/30 INDEFLATOR, INDEFLATOR Plus 30 and associated Priority Packs my exhibit leaks and/or a loose connection at the rotating luer assembly or stopcock connection, which could lead to air ingress under vacuum. Customers are instructed to: 1)Immediately stop using the devices from the affected lots 2) Review their inventory, complete and return the provided Effectiveness Check Form 3) Return all unused affected devices to the Recalling Firm 4) Share the Recall Notification with relevant personnel in their organization 5) if the affected products have been further distributed/transferred, notify those customers 6) Report any occurrence of product performance issues or patient adverse events to the Recalling Firm at 800-227-9902 The Recalling Firm: -Has immediately stop shipping devices from affected lots -Conducting an investigation to determine/confirm that there are not other affected products or lots in distribution -Implementing appropriate corrective actions to ensure product performance -Will work to replace inventory when available For questions or assistance contact the local representative or customer service department at 800-227-9902

Recall details

Recall number
Z-0925-2022
Status
Open, Classified
Initiated by the company
Mar 11, 2022
Recalling firm
Abbott Vascular
Quantity
7 lots
Product code
MAV
FDA submissions
K962495
Distribution
U.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Australia, Greece, R¿union, Belgium, Guam, Romania, Brazil, Indonesia, Russian Federation, Bahamas, Israel, Saudi Arabia, Canada, India, Singapore, Switzerland, Italy, Slovakia, Chile, Jamaica, Thailand, China, Japan, Taiwan Republic of China, Colombia, Korea (South), Costa Rica, Luxembourg, Uruguay, Czech Republic, Mexico, Vietnam, Germany, Netherlands, South Africa, Dominican Republic, New Zealand, Spain, and Panama

Product description

Plus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REF 1000185-115 WPL2122279-01 (2019-03-01)

Code information (models, lots, serial numbers)

Part Numbers: 1000185 Device Identifier-GTIN: 8717648013607 Lot Numbers: 60317532 60323314 60329935 60323313 60326601  Part Numbers: 1000185-115 Device Identifier-GTIN: 08717648015267 Lot Numbers: 60316763 60323315

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89760).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page