NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx… recallZ-1000-2022 · NeuMoDx Molecular Inc · initiated Feb 18, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from NeuMoDx Molecular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients
FDA root cause category: Material/Component Contamination
What the company is doing
QIAGEN issued customer Urgent Medical Device Correction notification via email firstly by English (then in local language if required)on 2/18/22. Letter states reason for recall, health risk and action to take: 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Field Safety Corrective Action with all users of the NeuMoDx Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDx Cartridges from any of these lots in combination with the NeuMoDx SARS-CoV-2 Assay, the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay and NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this Urgent Field Safety Corrective Action. 4. Complete the attached response form and return to [email protected] to confirm receipt of this notification. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or [email protected]
Recall details
- Recall number
- Z-1000-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 18, 2022
- Recalling firm
- NeuMoDx Molecular Inc
- Quantity
- 773 (72 US and 701 ROW) cases (48 cartridges/case)
- Product code
- QJR
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of OH, MO, VA, MI, MI, FL, MD, IA, NY, MO and the countries of Italy, Germany, Switzerland, Sweden, Netherlands, Finland, Spain, United Kingdom, Belgium, Austria, Hong Kong, France, Denmark, Slovenia.
Product description
NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets other than SARS-CoV-2 are not impacted by this Urgent Medical Device Correction Ref. Number: 100100
Code information (models, lots, serial numbers)
Lot Numbers: 115424, 115425, 115426, 115427, 115429, 115431. GTIN: 10814278020274
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NeuMoDxTM Cartridge 100100Neumodx Molecular, Inc. · General purpose reagent3 recalls5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page