Software 1898072 IPC upgrade v 2.7.3.0, IPC System recallZ-0961-2022 · Medtronic Xomed, Inc. · initiated Dec 15, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Xomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rotations and the actual number of rotations in a specific mode when the product was used with our company's M5 handpiece.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm notified the Japanese consignees by letter on 12/12/2021. The letter stated the following: "In our company's records, the software version of your hospital's integrated power console (IPC) is not v2.7.3.0. However, we found that there were cases in other facilities that the software version was updated to v2.7.3.0 was not recorded in our company's records even though the software version was updated to v2.7.3.0.. Therefore, we proceed the following measures. Thank you for your understanding and cooperation. Content l If it has been confirmed that the software used in your hospital is v2.7.3.0, please contact your Medtronic representative or distributor representative and suspend to use with the M5 handpiece (the method of checking the software version is described in the separate sheet). l Medtronic representative or distributor representative contact and visit to check the software version of your hospital's IPC."

Recall details

Recall number
Z-0961-2022
Status
Terminated
Initiated by the company
Dec 15, 2021
Recalling firm
Medtronic Xomed, Inc.
Quantity
1447 units
Product code
ERL
FDA submissions
K081277, K081475
Distribution
International distribution country of Japan.

Product description

Software 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND EHS, Product Number EC300

Code information (models, lots, serial numbers)

Show the full code information (13k characters)

a) Product Number 1898001, GTIN 00613994752246, Lot Serial Numbers: 72158, 72162, 72167, 72178, 72176, 72205, 72206, 72208, 72234, 72323, 72315, 72326, 72339, 72340, 72342, 72372, 72404, 72419, 72422, 72471, 72509, 72492, 72788, 72818, 72819, 72815, 72816, 72838, 72843, 72639, 72607, 72630, 72632, 73117, 73114, 73115, 73122, 73204, 73206, 73218, 73554, 73354, 73355, 73358, 73359, 73351, 73384, 73408, 73410, 73416, 73420, 73421, 73427, 73431, 73432, 73433, 73430, 73434, 73508, 73510, 73507, 73509, 73539, 73540, 73560, 73561, 73604, 73830, 73844, 73847, 73848, 73851, 73852, 73853, 73855, 73846, 73859, 73856, 73863, 73864, 73874, 73875, 73877, 73878, 73879, 73992, 73993, 73994, 73995, 74014, 74017, 74016, 74036, 74033, 74046, 74047, 74048, 74050, 74055, 74057, 74061, 74062, 74227, 74229, 74231, 74232, 74310, 74313, 74314, 74315, 74317, 74353, 74354, 74386, 74385, 74394, 74434, 74436, 74512, 74577, 74571, 74587, 74590, 74592, 74594, 74595, 74596, 74646, 74647, 74644, 74674, 74763, 74766, 74770, 74769, 74773, 74854, 74880, 74881, 74899, 74900, 74978, 74979, 74980, 74982, 74988, 74989, 74991, 74992, 75008, 75144, 75166, 75206, 75226, 75272, 75313, 75315, 75317, 75422, 75423, 75424, 75426, 75428, 75430, 75456, 75459, 75460, 75462, 75492, 75493, 75495, 75494, 75526, 75529, 75543, 75708, 75709, 75710, 75711, 75712, 75713, 75714, 75725, 75726, 75727, 75769, 75848, 75849, 76056, 76090, 76129, 76127, 76145, 76169, 76171, 76334, 76335, 76360, 76363, 76365, 76362, 76413, 76414, 76476, 76492, 76537, 76589, 76592, 76667, 76668, 76669, 76742, 76744, 76745, 76746, 76747, 76748, 76749, 76797, 76801, 76810, 76811, ENTC77039, ENTC77041, ENTC77045, ENTC77047, ENTC77048, ENTC77122, ENTC77123, ENTC77138, ENTC77139, ENTC77141, ENTC77146, ENTC77180, ENTC77189, ENTC77194, ENTC77198, ENTC77199, ENTC77210, ENTC77496, ENTC77499, ENTC77500, ENTC77578, ENTC77584, ENTC77586, ENTC77587, ENTC77589, ENTC77591, ENTC77633, ENTC77697, ENTC77911, ENTC77912, ENTC77946, ENTC77949, ENTC77951, ENTC77993, ENTC77995, ENTC78052, ENTC78261, ENTC78262, ENTC78263, ENTC78264, ENTC78388, ENTC78390, ENTC78391, ENTC78393, ENTC78413, ENTC78453, ENTC78471, ENTC78472, ENTC78499, ENTC78577, ENTC78582, ENTC78596, ENTC78601, ENTC78659, ENTC78660, ENTC78683, ENTC78701, ENTC78704, ENTC78840, ENTC78842, ENTC78843, ENTC78983, ENTC78985, ENTC78986, ENTC78988, ENTC79028, ENTC79029, ENTC79030, ENTC78831, ENTC79060, ENTC79064, ENTC79088, ENTC79091, ENTC79092, ENTC79146, ENTC79147, ENTC79148, ENTC79149, ENTC79152, ENTC79154, ENTC79156, ENTC79157, ENTC79319, ENTC79361, ENTC79363, ENTC79364, ENTC79366, ENTC79367, ENTC79406, ENTC79409, ENTC79413, ENTC79417, ENTC79418, ENTC79419, ENTC79445, ENTC79449, ENTC79452, ENTC79467, ENTC79571, ENTC79598, ENTC79626, ENTC79627, ENTC79628, ENTC79629, ENTC79630, ENTC79633, ENTC79634, ENTC79637, ENTC79638, ENTC79656, ENTC79665, ENTC79666, ENTC79672, ENTC80074, ENTC80274, ENTC80276, ENTC80296, ENTC80480, ENTC80482, ENTC80483, ENTC80484, ENTC80506, ENTC80535, ENTC80551, ENTC80578, ENTC80582, ENTC80583, ENTC80625, ENTC80677, ENTC80678, ENTC80679, ENTC80680, ENTC80705, ENTC80707, ENTC80709, ENTC80710, ENTC80759, ENTC80816, ENTC80819, ENTC80837, ENTC80840, ENTC80873, ENTC80874, ENTC80875, ENTC80939, ENTC80941, ENTC80954, ENTC81013, ENTC81025, ENTC81031, ENTC81033, ENTC81035, ENTC81042, ENTC81044, ENTC81064, ENTC81070, ENTC81114, ENTC81116, ENTC81120, ENTC81121, ENTC81122, ENTC81123, ENTC81153, ENTC81262, ENTC81263, ENTC81458, ENTC81459, ENTC81496, ENTC81497, ENTC81499, ENTC81500, ENTC81505, ENTC81506, ENTC81507, ENTC81534, ENTC81535, ENTC81537, ENTC81615, ENTC81655, ENTC81657, ENTC81675, ENTC81690, ENTC81691, ENTC81692, ENTC81693, ENTC81694, ENTC81696, ENTC81697, ENTC81698, ENTC81782, ENTC81785, ENTC81786, ENTC81845, ENTC81846, ENTC81847, ENTC81849, ENTC81851, ENTC81852, ENTC82024, ENTC82027, ENTC82035, ENTC82036, ENTC82037, ENTC82083, ENTC82087, ENTC82090, ENTC82219, ENTC82223, ENTC82259, ENTC82263, ENTC82264, ENTC82275, ENTC82289, ENTC82290, ENTC82291, ENTC82294, ENTC82295, ENTC82302, ENTC82333, ENTC82664, ENTC82666, ENTC82687, ENTC82801, ENTC82806, ENTC82807, ENTC82809, ENTC82834, ENTC82835, ENTC82836, ENTC82856, ENTC82857, ENTC82862, ENTC82869, ENTC82873, ENTC82985, ENTC82987, ENTC82989, ENTC82992, ENTC82993, ENTC83174, ENTC83177, ENTC83181, ENTC83182, ENTC83185, ENTC83186, ENTC83187, ENTC83188, ENTC83190, ENTC83191, ENTC83208, ENTC83329, ENTC83333, ENTC83334, ENTC83370, ENTC83371, ENTC83372, ENTC83377, ENTC83378, ENTC83382, ENTC83384, ENTC83445, ENTC83446, ENTC83447, ENTC83715, ENTC83716, ENTC83722, ENTC83811, ENTC83814, ENTC83828, ENTC83829, ENTC83830, ENTC83831, ENTC83958, ENTC83964, ENTC83966, ENTC83969, ENTC83972, ENTC83973, ENTC84162, ENTC84173, ENTC84225, ENTC84271, ENTC84309, ENTC84314, ENTC84316, ENTC84320, ENTC84324, ENTC84406, ENTC84620, ENTC84623, ENTC84626 b) Product Number EC300RF, GTIN 00613994614629, Lot Serial Numbers: 51010, 51285, 51407, 51477, 51533, 51559, 51693, 51729, 52266, 52282, 52303, 52464, 52586, 53704, 53831, 54082, 54448, 54475, 54487, 54820, 55614, 57027 c) Product Number EC300, GTIN 00613994750167, Lot Serial Numbers: 52005, 52154, 52158, 52159, 52189, 52190, 52200, 52212, 52248, 52250, 52254, 52258, 52312, 52313, 52315, 52324, 52374, 52377, 52380, 52387, 52398, 52589, 52590, 52595, 52925, 53063, 53065, 53067, 53068, 53086, 53089, 53090, 53092, 53094, 53113, 53155, 53616, 53617, 53618, 53620, 53622, 53623, 53633, 53660, 53661, 53662, 53663, 53664, 53717, 53817, 53818, 53871, 53900, 53901, 53922, 53923, 53924, 53972, 53974, 53976, 53977, 53979, 53980, 53981, 53982, 53983, 53984, 53985, 53999, 54000, 54034, 54035, 54036, 54037, 54050, 54053, 54057, 54062, 54063, 54065, 54148, 54150, 54158, 54159, 54172, 54173, 54175, 54176, 54177, 54197, 54198, 54200, 54201, 54202, 54221, 54222, 54223, 54225, 54231, 54232, 54233, 54235, 54236, 54237, 54524, 54525, 54534, 54591, 54592, 54629, 54632, 54634, 54635, 54642, 54704, 54705, 54706, 54707, 54738, 54802, 54835, 54854, 54855, 54870, 54871, 54943, 54947, 54952, 54982, 54983, 55040, 55042, 55056, 55089, 55148, 55150, 55151, 55152, 55155, 55302, 55304, 55365, 55367, 55368, 55382, 55383, 55422, 55424, 55425, 55426, 55432, 55540, 55541, 55543, 55544, 55545, 55546, 55568, 55569, 55571, 55572, 55573, 55576, 55580, 55582, 55687, 55699, 55747, 55748, 55749, 55782, 55852, 55854, 55855, 56021, 56022, 56024, 56025, 56026, 56027, 56028, 56029, 56030, 56055, 56086, 56089, 56092, 56093, 56094, 56095, 56096, 56097, 56357, 56371, 56372, 56379, 56380, 56382, 56394, 56395, 56398, 56487, 56488, 56489, 56629, 56632, 56645, 56646, 56647, 56720, 56722, 56723, 56724, 56744, 56747, 56748, 56749, 56771, 56802, 56813, 56817, 56818, 56819, 56820, 56821, 56859, 56860, 56861, 56867, 56901, 56902, 56909, 56910, 56911, 56914, 56915, 57318, 57319, 57320, 57321, 57324, 57325, 57326, 57327, 57328, 57329, 57331, 57521, 57522, 57525, 57526, 57527, 57551, 57555, 57566, 57572, 57574, 57575, 57576, 57579, 57582, 57600, 57604, 57605, 57635, 57636, 57637, 57644, 57844, 57846, 57848, 57933, 57935, 57937, 57939, 57941, 57942, 57943, 58164, 58165, 58166, 58167, 58169, 58220, 58222, 58223, 58224, 58321, 58466, 58470, 58554, 58556, 58559, 58577, 58578, 58584, 58585, 58606, 58640, 58641, 58642, 58643, 58644, 58646, 58647, 58649, 58651, 58652, 58654, 58655, 58659, 58660, 58803, 58804, 58805, 58806, 58842, 58843, NTC59090, NTC59104, NTC59106, NTC59107, NTC59127, NTC59154, NTC59155, NTC59163, NTC59164, NTC59165, NTC59166, NTC59173, NTC59177, NTC59222, NTC59223, NTC59225, NTC59228, NTC59230, NTC59231, NTC59320, NTC59322, NTC59323, NTC59337, NTC59339, NTC59349, NTC59351, NTC59355, NTC59418, NTC59420, NTC59421, NTC59422, NTC59423, NTC59425, NTC59466, NTC59468, NTC59469, NTC59470, NTC59472, NTC59479, NTC59498, NTC59499, NTC59501, NTC59567, NTC59568, NTC59569, NTC59570, NTC59571, NTC59572, NTC59573, NTC59576, NTC59577, NTC59843, NTC59844, NTC59892, NTC59896, NTC59914, NTC59929, NTC59931, NTC59932, NTC59935, NTC59937, NTC59959, NTC59963, NTC59967, NTC59973, NTC59987, NTC59991, NTC59993, NTC60022, NTC60030, NTC60031, NTC60033, NTC60036, NTC60043, NTC60067, NTC60074, NTC60114, NTC60122, NTC60127, NTC60169, NTC60181, NTC60185, NTC60190, NTC60237, NTC60239, NTC60266, NTC60310, NTC60311, NTC60312, NTC60421, NTC60422, NTC60466, NTC60469, NTC60472, NTC60473, NTC60478, NTC60539, NTC60564, NTC60572, NTC60574, NTC60575, NTC60578, NTC60608, NTC60726, NTC60730, NTC60751, NTC60770, NTC60774, NTC60776, NTC60780, NTC60847, NTC60857, NTC60858, NTC60860, NTC60866, NTC60904, NTC60909, NTC60910, NTC60911, NTC60939, NTC60940, NTC60941, NTC60942, NTC60943, NTC61107, NTC61121, NTC61259, NTC61275, NTC61328, NTC61331, NTC61332, NTC61333, NTC61334, NTC61335, NTC61336, NTC61337, NTC61338, NTC61339, NTC61340, NTC61341, NTC61343, NTC61344, NTC61345, NTC61578, NTC61593, NTC61622, NTC61623, NTC61624, NTC61628, NTC61630, NTC61631, NTC61633, NTC61634, NTC61635, NTC61636, NTC61637, NTC61638, NTC61639, NTC61659, NTC61660, NTC61826, NTC61836, NTC61864, NTC61867, NTC61868, NTC61934, NTC61935, NTC61936, NTC61937, NTC61938, NTC61941, NTC61981, NTC62063, NTC62074, NTC62087, NTC62088, NTC62089, NTC62090, NTC62091, NTC62092, NTC62093, NTC62094, NTC62095, NTC62097, NTC62243, NTC62244, NTC62246, NTC62247, NTC62248, NTC62249, NTC62251, NTC62252, NTC62254, NTC62255, NTC62256, NTC62258, NTC62259, NTC62260, NTC62261, NTC62262, NTC62264, NTC62266, NTC62268, NTC62269, NTC62270, NTC62271, NTC62274, NTC62275, NTC62276, NTC62495, NTC62519, NTC62520, NTC62521, NTC62526, NTC62530, NTC62534, NTC62541, NTC62545, NTC62583, NTC62588, NTC62593, NTC62594, NTC62596, NTC62597, NTC62599, NTC62601, NTC62655, NTC62656, NTC62657, NTC62658, NTC62659, NTC62676, NTC62679, NTC62684, NTC62685, NTC62686, NTC62687, NTC62689, NTC62691, NTC63140, NTC63181, NTC63182, NTC63183, NTC63184, NTC63185, NTC63226, NTC63227, NTC63230, NTC63231, NTC63232, NTC63234, NTC63235, NTC63545, NTC63574, NTC63575, NTC63576, NTC63577, NTC63733, NTC63755, NTC63787, NTC63790, NTC63793, NTC63794, NTC63795, NTC63796, NTC64148, NTC64149, NTC64194, NTC64239, NTC64286, NTC64323, NTC64330, NTC64333, NTC64335, NTC64336, NTC64377, NTC64431, NTC64441, NTC64453, NTC64455, NTC64456, NTC64457, NTC64458, NTC64459, NTC64460, NTC64461, NTC64462, NTC64464, NTC64465, NTC64466, NTC64469, NTC64470, NTC64473, NTC64474, NTC64478, NTC64480, NTC64481, NTC64482, NTC64483, NTC64484, NTC64485, NTC64487, NTC64509, NTC64512, NTC64515, NTC64516, NTC64519, NTC64523, NTC64524, NTC64525, NTC64526, NTC64527, NTC64528, NTC64529, NTC64530, NTC64531, NTC64532, NTC64577, NTC64644, NTC64898, NTC64900, NTC64901, NTC64903, NTC64905, NTC65008, NTC65009, NTC65010, NTC65011, NTC65012, NTC65013, NTC65014, NTC65015, NTC65016, NTC65046, NTC65047, NTC65048, NTC65168, NTC65169, NTC65261, NTC65262, NTC65263, NTC65264, NTC65265, NTC65266, NTC65268, NTC65270, NTC65271, NTC65272, NTC65781, NTC65785, NTC65789, NTC65799, NTC65801, NTC65802, NTC65817, NTC65818, NTC65850, NTC65859, NTC65960, NTC65963, NTC65964, NTC65965, NTC65987, NTC65988, NTC65989, NTC65991, NTC65992, NTC65995, NTC66046, NTC66047, NTC66048, NTC66049, NTC66051, NTC66053, NTC66057, NTC66058, NTC66059, NTC66060, NTC66061, NTC66062, NTC66063, NTC66064, NTC66065, NTC66066, NTC66068, NTC66069, NTC66070, NTC66071, NTC66072, NTC66073, NTC66075, NTC66079, NTC66080, NTC66081, NTC66082, NTC66086, NTC66087, NTC66088, NTC66089, NTC66100, NTC66101, NTC66111, NTC66112, NTC66120, NTC66138, NTC66139, NTC66140, NTC66141, NTC66142, NTC66143, NTC66149, NTC66188, NTC66189, NTC66190, NTC66191, NTC66192, NTC66193, NTC66194, NTC66212, NTC66213, NTC66214, NTC66215, NTC66216, NTC66217, NTC66218, NTC66219, NTC66243, NTC66282, NTC66283, NTC66284, NTC66285, NTC66286, NTC66872, NTC66925, NTC66926, NTC66927, NTC66928, NTC66929, NTC66930, NTC66931, NTC66939, NTC66941, NTC66942, NTC66944, NTC66993, NTC66994, NTC66995, NTC66996, NTC66997, NTC66998, NTC66999, NTC67000, NTC67001, NTC67002, NTC67003, NTC67004, NTC67005, NTC67226, NTC67305, NTC67307, NTC67347, NTC67348, NTC67350, NTC67351, NTC67352, NTC67354, NTC67355, NTC67356, NTC67358, NTC67359, NTC67360, NTC67361, NTC67363, NTC67365, NTC67366, NTC67367, NTC67369, NTC67370, NTC67371, NTC67372, NTC67373, NTC67438, NTC67447, NTC67448, NTC67449, NTC67465, NTC67468, NTC67470, NTC67471, NTC67472, NTC67473, NTC67476, NTC67477, NTC67478, NTC67480, NTC67481, NTC67482, NTC67497, NTC67499, NTC67500, NTC67501, NTC67503, NTC67504, NTC67505, NTC67620, NTC67705, NTC67716, NTC67727, NTC67804, NTC67808, NTC68018, NTC68019, NTC68022, NTC68152, NTC68155, NTC68156, NTC68157, NTC68158, NTC68159, NTC68160, NTC68161, NTC68163, NTC68166, NTC68167, NTC68168, NTC68169, NTC68215, NTC68216, NTC68217, NTC68218, NTC68219, NTC68226, NTC68227, NTC68228, NTC68229, NTC68230, NTC68232, NTC68237, NTC68241, NTC68242, NTC68243, NTC68245, NTC68250, NTC68253, NTC68254, NTC68255, NTC68256, NTC68258, NTC68306, NTC68308, NTC68442, NTC68470, NTC68481, NTC68486, NTC68645, NTC68646, NTC68649, NTC68650, NTC68651, NTC68652, NTC68653, NTC68754, NTC68755, NTC68756, NTC68757, NTC69003, NTC69004, NTC69239, NTC69242, NTC69243, NTC69250, NTC69251, NTC69252, NTC69281, NTC69320, NTC69321, NTC69322, NTC69328, NTC69329, NTC69330, NTC69331, NTC69332, NTC69333, NTC69334, NTC69335, NTC69336, NTC69353, NTC69354, NTC69454, NTC69513, NTC69588, NTC69800, NTC71502

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

47 devices cleared under K081277, K081475 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page