SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to… recallZ-1003-2022 · Boston Scientific Corporation · initiated Feb 24, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
FDA root cause category: Labeling design
What the company is doing
An URGENT MEDICAL DEVICE PRODUCT ADVISORY dated 2/24/22 was sent to customers. Inadvertent placement of SpaceOAR gel into a blood vessel and subsequent migration of the hydrogel outside of the pelvis, led to embolism. Boston Scientific will incorporate best practices regarding hydrogel placement during all SpaceOAR procedures into the IFU. The updated IFUs will be packaged with SpaceOAR and SpaceOAR Vue Systems after regulatory approvals are obtained for the IFU updates. This Product Advisory contains the recommended IFU updates, located in Appendix 1, intended to: -Raise awareness of embolism as an existing adverse event. -Add new Warnings and Precautions that describe technique recommendations for the placement of SpaceOAR and SpaceOAR Vue to reduce the unintentional placement of hydrogel. - Add procedural instructions to reduce the unintentional placement of hydrogel in a blood vessel and steps to verify placement of the hydrogel following implantation. NO product is being recalled and you are NOT required to return product to Boston Scientific. As per the IFU, SpaceOAR and SpaceOAR Vue System procedures should only be performed by physicians who have received appropriate training for proper hydrogel spacer deployment technique. The Acknowledgement Form must be completed and returned even if you do not have any affected units. Return the Acknowledgement Form by Email: [email protected] or Fax to: Field Action Center 1-763-415-7708
Recall details
- Recall number
- Z-1003-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 24, 2022
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 23,368 units
- Product code
- OVB
- FDA submissions
- K182971
- Distribution
- Worldwide - US Nationwide distribution.
Product description
SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
UPN SV-1010, GTIN 00850009803009, Batch Numbers: 26383209, 26383210, 26517604, 26861584, 26861585, 26891612, 26891613, 27083784, 27190242, 27210560, 27257801, 27257802, 27455966, 27455967, 27496959, 27497140, 27670956, 27708041, 27708042, 27708043, 27930863, 27930864, 27941129, 27941130, 27949315, 28094077, 28247077, 28247078, 28358778, 28377018, 28377019, 28464794, 28464795, 28496461, 28496462, 28578758, 28578759, 28608231, 28608232; UPN SV-2101, GTIN 00864661000140, Batch Numbers: 26345481, 26376961, 26376962, 26376963, 26376964, 26380456, 26380457, 26384055, 26384056, 26408849, 26485974, 26485975, 26485977, 26486586, 26486587, 26486588, 26493461, 26517606, 26517607, 26529719, 26567283, 26567284, 26567285, 26567286, 26567287, 26574373, 26574374, 26574378, 26574379, 26574441, 26611593, 26611594, 26611595, 26652286, 26652287, 26691824, 26691826, 26691827, 26691828, 26705563, 26705564, 26705565, 26705566, 26728564, 26728565, 26743769, 26743770, 26743771, 26743772, 26745057, 26770161, 26770162, 26770163, 26770164, 26770165, 26861586, 26880769, 26880770, 26880771, 26880772, 26898042, 26898043, 26898044, 26898045, 26899812, 26899813, 26906603, 26906604, 26929712, 26929713, 26929714, 26929715, 26929716, 26929717, 26929719, 26929860, 26929861, 26957514, 26957515, 26957516, 26957517, 26969537, 26969538, 26969539, 26970440, 26970441, 26970442, 27010832, 27040235, 27040236, 27060422, 27060423, 27060424, 27060425, 27060426, 27078872, 27078873, 27084457, 27084458, 27084459, 27084960, 27100768, 27100769, 27100770, 27123402, 27123404, 27140485, 27140486, 27163487, 27163488, 27180605, 27190243, 27190244, 27190245, 27190686, 27193837, 27193838, 27193839, 27194460, 27194461, 27197622, 27197623, 27214381, 27214382, 27214383, 27224419, 27257804, 27257805, 27259062, 27259063, 27259064, 27259065, 27268186, 27268187, 27283462, 27283463, 27283464, 27283465, 27283466, 27317224, 27317225, 27317226, 27329273, 27329274, 27329275, 27358004, 27358005, 27361380, 27361381, 27382203, 27382204, 27416003, 27416004, 27416005, 27416006, 27416007, 27444898, 27444899, 27445360, 27447905, 27469811, 27469812, 27469813, 27475230, 27496957, 27496958, 27512052, 27512053, 27512054, 27512055, 27517685, 27518868, 27518869, 27528492, 27531152, 27531153, 27531154, 27531155, 27531156, 27531157, 27601494, 27601495, 27657444, 27657445, 27657446, 27670942, 27670945, 27670946, 27670949, 27705891, 27720117, 27720118, 27720119, 27745215, 27745216, 27756475, 27769338, 27769339, 27769860, 27769861, 27769862, 27769863, 27797083, 27797084, 27797085, 27797086, 27797087, 28094078, 28094079, 28094320, 28103330, 28104177, 28104178, 28104179, 28104420, 28110684, 28110685, 28110686, 28110687, 28116634, 28116635, 28138219, 28138440, 28146630, 28146631, 28169753, 28169754, 28169755, 28179669, 28179670, 28179671, 28202160, 28237387, 28237388, 28237389, 28247079, 28247160, 28247161, 28247162, 28247269, 28247270, 28247271, 28247272, 28254386, 28254387, 28254388, 28265943, 28265944, 28265945, 28265946, 28265947, 28272556, 28272557, 28272558, 28272559, 28272560, 28287355, 28287356, 28294718, 28294719, 28311111, 28311112, 28323414, 28323415, 28323416, 28323417, 28335613, 28335614, 28350366, 28350367, 28350368, 28350369, 28364006, 28364007, 28364008, 28371227, 28371228, 28371229, 28389635, 28389636, 28412353, 28419102, 28419103, 28432892, 28432893, 28473289, 28473290, 28493777, 28493778, 28499505, 28499506, 28499507, 28555843, 28565355, 28565356, 28565357, 28587389, 28587390, 28587391, 28587392, 28587393, 28587394, 28605867, 28606252, 28611099, 28612440, 28612441, 28612442, 28613189, 28635760, 28635761, 28635762, 28635763, 28654059, 28654440, 28654441, 28663263, 28663264, 28663265, 28668946, 28668947, 28668948, 28683114, 28683115, 28686271, 28686272, 28686273, 28694687, 28707785, 28709642, 28709643, 28714572, 28723144, 28734763, 28734764, 28758266, 28766992, 28766993, 28766994, 28766995, 28766996, 28811555, 28811556, 28811558, 28811560, 28819972, 28819973, 28821692, 28821693, 28822526, 28822527, 28828281, 28828282, 28836638, 28836639, 28838125
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SpaceOAR VUE™ System - 10mL SV-2101Boston Scientific Corporation · Hydrogel spacer1 recall4 deaths414 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89862).
- Z-1002-2022 · Open, ClassifiedSpaceOAR System, UPN SO-1010, SO-2101, SO-3101, and SO-4101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer t…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page