SMART TOE recallZ-1011-2022 · Stryker GmbH · initiated Mar 28, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The nickel-sensitivity statement on the outer carton may be partially covered by a label resulting in the statement showing "Indicated for patients with Nickel sensitivity" instead of "Not Indicated for patients with Nickel sensitivity".

FDA root cause category: Process control

What the company is doing

A recall notice was disseminated via mail on 03/28/2022 delivered by FedEx Priority One to Hospital/Medical Facilities and Distributor/Subsidiary via email where inventory records indicate the product was distributed. They are requested to immediately quarantine all recalled products they have and return all affected product to local Stryker GmbH distribution site as soon as possible. If they have further distributed the affected product, they are asked to notify the applicable parties at once about this recall and Stryker will work directly with them to collect the recalled items.

Recall details

Recall number
Z-1011-2022
Status
Open, Classified
Initiated by the company
Mar 28, 2022
Recalling firm
Stryker GmbH
Quantity
246 implants
Product code
HTY
FDA submissions
K112197
Distribution
Worldwide distribution - US Nationwide and the countries including Australia, Canada, Chile, China, France, Germany, Netherlands, New Zealand, Spain, Sweden, Switzerland, and United Kingdom.

Product description

Smart Toe II Intramedullary Arthrodesis Implant

Code information (models, lots, serial numbers)

Model Number ST0XS-13; UDI: 07613252263470; Lot Code (Expiration Date): H52262 (08/31/2026) Model Number ST0-19P; UDI: 07613252263340; Lot Code (Expiration Date): H54191 (08/31/2026) Model Number ST0A-16P; UDI: 07613252263418; Lot Code (Expiration Date): H54193 (09/30/2026) Model Number ST0XS-11; UDI: 07613252263456; Lot Code (Expiration Date): H55690 (09/30/2026) Model Number ST0-22P; UDI: 07613252263388; Lot Code (Expiration Date): H56502 (09/30/2026)

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates