ACS IR - Angio Pack recallZ-1038-2022 · American Contract Systems, Inc. · initiated Mar 22, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Contract Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product was sterilized with a higher than specification EO concentration.

FDA root cause category: Process control

What the company is doing

A customer letter dated 04/05/2022 was issued to impacted consignees. The letter instructs the following: 1) Review your inventory and segregate affected product. This product should not be shipped out to customers. 2) Return affected product to 2610 NE Industrial Drive, Suite 220, Kansas City, MO 64120. 3) If product was already further distributed, ensure that all downstream customers are properly notified of the recall and maintain records of effectiveness. A Field Action Response Form should be completed and returned to [email protected], even if you do not have product inventory.

Recall details

Recall number
Z-1038-2022
Status
Open, Classified
Initiated by the company
Mar 22, 2022
Recalling firm
American Contract Systems, Inc.
Quantity
24 packs
Product code
OIP
Distribution
US Nationwide distribution in the states of IL, MO, NE, and TX.

Product description

ACS IR - Angio Pack

Code information (models, lots, serial numbers)

Model JEIR37H, Lot 657211 UDI: 00191072151612

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 89868).

Sources and updates