Philips Hemodynamic Application, Model Number 722463… recallZ-1041-2022 · Philips North America Llc · initiated Feb 22, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT Medical Device Correction notification letter dated 2/18/22 was sent to customers. Actions that should be taken by the customer/ user in order to prevent risks for patients. Do not use the Non-Invasive Blood Pressure for Neonatal Patients until Philips has installed a new software revision that will solve this problem. Place this Urgent Medical Device Correction letter with the documentation of the Philips Hemodynamic Application until Philips implements the new software revision. Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Urgent Medical Device Correction letter. Actions planned by Philips to correct the problem. To correct this problem, Philips will install a new software revision to update the Hemodynamic Application to revision Rl.3. This new revision will be available as of March 2022. Philips will be contacting customers with an affected Philips Hemodynamic Application to install the new software revision. Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information, please contact your local Philips Cardiology representative.
Recall details
- Recall number
- Z-1041-2022
- Status
- Terminated
- Initiated by the company
- Feb 22, 2022
- Recalling firm
- Philips North America Llc
- Quantity
- 77
- Product code
- MWI
- FDA submissions
- K181311
- Distribution
- Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.
Product description
Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.
Code information (models, lots, serial numbers)
UDI: B-00884838084544; System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 886 14 15 643 1194 220 1238 947 949 950 951 89 657 169 168 413 414 624 565 1060 393 512 278 818 362 123 802 395 576 932 291 834 539 605 161 970 686 761 229 230
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Philips Hemodynamic Application Philips Hemodynamic Application - 1Philips Medical Systems Nederland B.V. · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 recall3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page