C-VIEW recallZ-1062-2022 · Karl Storz Endoscopy · initiated Apr 1, 2022
Class II Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Karl Storz Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
FDA root cause category: Under Investigation by firm
What the company is doing
On 04/01 and 6/2022, the firm emailed an "Urgent Medical Device Recall" letter to customers and followed up with a hardcopy letter on 04/07-08/2022 via their Third Party contractor informing customers of a labeling update to correct the instructions for use for certain flexible urological endoscopes and advise that high-level disinfection should not be used as a method of reprocessing. Supplemental validation testing of the efficacy of the manual high-level disinfection process using CIDEX OPA was performed. Testing showed that required efficacy level of disinfection using CIDEX OPA was not achieved. To the extent included in the current applicable instructions for use, the following methods are being removed as reprocessing methods from the instructions for use of the affected endoscopes. +Manual High-Level Disinfection: -Cidex-OPA -Revital-Ox" RESERT (e.g., 2.0% Accelerated Hydrogen Peroxide solution). +Automated Endoscope Reprocessor: -Medivators AER -Reliance EPS -EvoTech ECR Customer Actions: 1. Immediately discontinue use of all high-level disinfection methods for reprocessing affected endoscopes. 2. The affected endoscopes should be sterilized after each use by one of the sterilization methods recommended in the instructions for use specific to each endoscope. 3. If your facility does not have access to a sterilization method recommended in the applicable instructions for use, immediately discontinue use of the affected endoscopes and return products to KARL STORZ. 4. Ensure all relevant personnel in your organization are aware of recall. 5. Updated instructions for use will be available at https://spwebspace.karlstorz.com/sites/HLDCorrection/SitePages/Home.aspx . Once the updated instructions for use are available, discard any prior versions of the instructions for use. Questions: Telephone: 1-888-352-9616 Fax: 1-888-912-7088 E-Mail: [email protected]
Recall details
- Recall number
- Z-1062-2022
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 1, 2022
- Recalling firm
- Karl Storz Endoscopy
- Quantity
- All Serial Numbers manufactured/distributed since January 2018
- Product code
- FBO
- FDA submissions
- K131364
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY. O.U.S.: Mexico, Lebanon, and South Korea
Product description
11272V CMOS VIDEO-CYSTO-URETHROSCOPE REV 11272VA 11272VAK CMOS VIDEO-CYSTO-URETHROSCOPE REV; 11272VU 11272VUK CMOS VIDEO-CYSTO-URETHROSCOPE US; 11272VUA 11272VUAK CMOS VIDEO-CYSTO-URETHROSCOPE US
Code information (models, lots, serial numbers)
11272V N/A Flexible CMOS Video Cysto Urethroscope Z18446US-BE (01/2020) UDI Code: 4048551313350; 11272VA 11272VAK Flexible CMOS Video Cysto Urethroscope Z18446US-BE (01/2020) UDI Code: 4048551402481; 11272VU 11272VUK Flexible CMOS Video Cysto Urethroscope Z18446US-BE (01/2020) UDI Code: 4048551313367; 11272VUA 11272VUAK Flexible CMOS Video Cysto Urethroscope Z18446US-BE (01/2020) UDI Code: 4048551402559; All Serial Numbers manufactured/distributed since January 2018.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- C-VIEW 11272VKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VUKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VUAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 89880).
- Z-1061-2022 · Open, ClassifiedCYSTO-URETHRO-FIBERSCOPE REF 11272C1 CYSTOSCOPE, 15FR X 370MM, ADULT REF 11272C2 and Kit Number 11272CK2 CYSTO-URETHRO-FIBERSCOPE REF 11272CU1 and Kit Number 11272CUK1
- Z-1064-2022 · Open, Classified11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018); 11272VNU 11272VNUK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018)
- Z-1065-2022 · Open, Classified11272VUE 11272VUEK Flexible Video Cysto-Urethroscope 96136031USCA V1.1 (04/2021); 11272VUE-R 11272VUEK-R Flexible Video Cysto-Urethroscope 96136031USCA V1.1 (04/2021)
- Z-1067-2022 · Open, Classified11278A2 11278AK2 Flexible Ureteroscope Z20615US-BA (08-2018); 11278AC1 N/A Flexible Ureteroscope Z17859US-A (04/2016); 11278AC2 11278ACK2 Flexible Ureteroscope Z20615US-BA (08-2018…
- Z-1068-2022 · Open, Classified11278V 11278VK CMOS Video Ureteroscope Z17744US-A (04/2016); 11278VA 11278VAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VU 11278VUK Flexible Video-Uret…
- Z-1069-2022 · Open, Classified11278VSA 11278VSAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSE 11278VSEK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSU 1…
- Z-1070-2022 · Open, Classified11278VSA 11278VSAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSE 11278VSEK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSU 1…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.