11005BC1 11005BCK1 Flexible Bronchoscope Z21493US-BA… recallZ-1055-2022 · Karl Storz Endoscopy · initiated Apr 1, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Karl Storz Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
FDA root cause category: Under Investigation by firm
What the company is doing
On 04/01 and 6/2022, the firm emailed an "Urgent Medical Device Recall" letter to customers and followed up with a hardcopy letter on 04/07-08/2022 via their Third Party contractor informing customers of a labeling update to correct the instructions for use for certain flexible urological endoscopes and advise that high-level disinfection should not be used as a method of reprocessing. Supplemental validation testing of the efficacy of the manual high-level disinfection process using CIDEX OPA was performed. Testing showed that the required efficacy level of disinfection using CIDEX OPA was not achieved. To the extent included in the current applicable instructions for use, the following methods are being removed as reprocessing methods from the instructions for use of the affected endoscopes. +Manual High-Level Disinfection: -Cidex-OPA -Revital-Ox" RESERT (e.g., 2.0% Accelerated Hydrogen Peroxide solution). +Automated Endoscope Reprocessor: -Medivators AER -Reliance EPS -EvoTech ECR Customer Actions: 1. Immediately discontinue the use of all high-level disinfection methods for reprocessing the affected endoscopes. 2. The affected endoscopes should be sterilized after each use by one of the sterilization methods recommended in the instructions for use specific to each endoscope. 3. If your facility does not have access to a sterilization method recommended in the applicable instructions for use, immediately discontinue use of the affected endoscopes and return products to KARL STORZ. 4. Ensure that all relevant personnel in your organization are aware of recall. 5. Updated instructions for use will be available at https://spwebspace.karlstorz.com/sites/HLDCorrection/SitePages/Home.aspx . Once the updated instructions for use are available, discard any prior versions of the instructions for use. Any questions: Telephone: 1-888-352-9616 Fax: 1-888-912-7088 E-Mail: [email protected]
Recall details
- Recall number
- Z-1055-2022
- Status
- Open, Classified
- Initiated by the company
- Apr 1, 2022
- Recalling firm
- Karl Storz Endoscopy
- Quantity
- All Serial Numbers manufactured/distributed since January 2018
- Product code
- EOQ
- FDA submissions
- K212656
- Distribution
- U.S. Nationwide distribution in the states of FL, KY, LA, MA, MI, MO, MT, NC, NH, NJ, OK, SC, TX and WA.
Product description
11005BC1 11005BCK1 Flexible Bronchoscope Z21493US-BA (06/2019)
Code information (models, lots, serial numbers)
11005BC1 11005BCK1 Flexible Bronchoscope Z21493US-BA (06/2019) UDI Code: 04048551225738 All Serial Numbers manufactured/distributed since January 2018
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- n.a. 11005BC1Karl Storz GmbH & Co. KG · BRONCHOSCOPE (FLEXIBLE OR RIGID)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page