IN.PACT AV recallZ-1043-2022 · Medtronic Vascular, Inc. · initiated Mar 23, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Packaging for Balloon Catheters may be damaged resulting in loss of sterility.
FDA root cause category: Packaging process control
What the company is doing
On March 23, 2022 Medtronic issued a "Urgent: Medical Device Recall" notification to affected consignees. In addition to informing consignees about the recalled product, Medtronic ask consignees to take the following actions: 1. Identify and quarantine all unused affected IN.PACT Admiral and IN.PACT AV catheters as listed in Table 1. 2. Return all unused affected product in your inventory to Medtronic. Contact Medtronic Vascular Customer Service at 800-716-6700 to initiate a product return. Your local Medtronic Representative can assist you with the initiation of the return. 3. Please complete the enclosed Customer Confirmation Form and email to [email protected]. 4. This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.
Recall details
- Recall number
- Z-1043-2022
- Status
- Open, Classified
- Initiated by the company
- Mar 23, 2022
- Recalling firm
- Medtronic Vascular, Inc.
- Quantity
- 3 units
- Product code
- PRC
- FDA submissions
- P190008
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, PUERTO RICO, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Croatia Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Republic Of Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Oman, Peru, Philippines, Poland, Portugal, Puerto Rico, Reunion, Russian Federation, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom.
Product description
Medtronic IN.PACT AV, Paclitaxel-coated PTA Balloon Catheter, REF: IAV06008008P, Sterile EO, Rx Only
Code information (models, lots, serial numbers)
Lot Number: 0010990608/ GTIN: 00763000232580
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IN.PACT™ AV IAV06008008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter1 recall1 report
Product lines
Other devices under the same FDA submission
17 devices cleared under P190008 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- IN.PACT™ AV IAV08004008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter21 reports
- IN.PACT™ AV IAV08006008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter19 reports
- IN.PACT™ AV IAV08008008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter11 reports
- IN.PACT™ AV IAV09008008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter11 reports
- IN.PACT™ AV IAV12004008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter9 reports
- IN.PACT™ AV IAV06006008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter8 reports
- IN.PACT™ AV IAV09004008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter8 reports
- IN.PACT™ AV IAV06004004PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter1 death8 reports
- IN.PACT™ AV IAV07004008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter6 reports
- IN.PACT™ AV IAV10004008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter6 reports
- IN.PACT™ AV IAV07004004PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter5 reports
- IN.PACT™ AV IAV05008008PMedtronic, Inc. · Drug-coated peripheral transluminal angioplasty catheter1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 89935).
- Z-1042-2022 · Open, ClassifiedMedtronic IN.PACT Admiral, Paclitaxel-coated PTA Balloon Catheter, Sterile EO, Rx Only for the following model Numbers:ADM04004013P ADM04008013P ADM04008013P ADM04008013P ADM040120…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page