Skip to content

OnGuard2 CSTD Bag Adaptor SP A Chemfort CSTD Product… recallZ-1057-2022 · Simplivia Healthcare Ltd. · initiated Mar 8, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Simplivia Healthcare Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Disconnection of the OnGuard2 Bag Adaptor SP from infusion bags which may result in spillage of hazardous medications.

FDA root cause category: Equipment maintenance

What the company is doing

The recall notice and acknowledgement form dated 08 March 2022 were sent under B. Braun cover letter on 03/10/2022 via Certified USPS mail. Distributors are to forward the recall notification to customers to extend the recall to the end user/customer level. Identify current inventory, cease use and quarantine product subject to recall. Do not destroy, return product and complete the Product Removal Acknowledgement form. If you are a direct consignee of B. Braun Medical Inc. a B. Braun customer service representative will contact you to arrange for return of product. If you have received this recall notice from a customer of B. Braun, please return product to your local distributor to ensure proper credit and replacement of material. OnGuard Spike Port Adaptor may be used as an alternative to replace the OnGuard2 Bag Adaptor SP. If you have questions contact Simplivia: Shay Shaham, Sunday to Friday, 08:00 to 19:00 (GMT+2) at the following phone number: 97246908826.

Recall details

Recall number
Z-1057-2022
Status
Open, Classified
Initiated by the company
Mar 8, 2022
Recalling firm
Simplivia Healthcare Ltd.
Quantity
293,520
Product code
ONB
FDA submissions
K192866
Distribution
Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.

Product description

OnGuard2 CSTD Bag Adaptor SP A Chemfort CSTD Product, Product code MG412143

Code information (models, lots, serial numbers)

UDI: 37290115921435, Lots: UBE507, UBE508, UBE513, UBE516, UBE517, UBE509, UBE511, UBE512, UBE514, UBE515, UBE517, UBE518, UBE526, UBE530, UBE531, UBE534, UBE533, UBE535, UBE536

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K192866 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page