ADVIA Centaur eE2 recallZ-1023-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Mar 3, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued an Urgent Medical Device Correction AIMC 22-03.A.US and Urgent Field Safety Notice AIMC 22-03.A.OUS on 3/3/22. Letter states reason, health risk and action to take: Customers must discontinue use of plasma specimens for testing with the Atellica IM and ADVIA Centaur eE2 assays until further notice. Customers may continue to use the Atellica IM eE2 and ADVIA Centaur eE2 assays with serum specimens. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Questions: contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative
Recall details
- Recall number
- Z-1023-2022
- Status
- Open, Classified
- Initiated by the company
- Mar 3, 2022
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 2,315 units
- Product code
- CHP
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Bulgaria, Burkina, Faso Country AG, Cyprus, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Greece, Guadeloupe, India, Ireland, Israel, Italy, Kazakhstan, Latvia, Lithuania, Macedonia, Malta, Morocco, Nepal, Netherlands, Norway, Palestinian Ter, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Vatican city.
Product description
ADVIA Centaur Enhanced Estradiol (eE2) 100T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10490889 All lots
Code information (models, lots, serial numbers)
All lot numbers UDI: (01)00630414573342(10)08930076(17)20220115 (01)00630414573342(10)10982085(17)20221108 (01)00630414573342(10)19929072(17)20210819 (01)00630414573342(10)20298079(17)20220304 (01)00630414573342(10)48171081(17)20220304 (01)00630414573342(10)53554081(17)20220521 (01)00630414573342(10)70551083(17)20220805 (01)00630414573342(10)76055076(17)20220115
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA Centaur® eE2 10490889Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, ESTRADIOL1 recall8 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 89953).
- Z-1020-2022 · Open, ClassifiedAtellica IM Enhanced Estradiol (eE2) 100T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10995561 All lots
- Z-1021-2022 · Open, ClassifiedAtellica IM Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10995562 All lots
- Z-1022-2022 · Open, ClassifiedAtellica IM Enhanced Estradiol (eE2) (REF) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 11206250 All lots
- Z-1024-2022 · Open, ClassifiedADVIA Centaur Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10491445 All lots
- Z-1025-2022 · Open, ClassifiedADVIA Centaur Enhanced Estradiol (eE2) (REF) 500T - In vitro diagnostic use in the quantitative detection of estradiol SMN: 10697757 All lots
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page