JOURNEY recallZ-1099-2022 · Smith & Nephew Inc · initiated Apr 19, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was reported that a JOURNEY II BCS Femoral OXIN LT SZ5 was in a box labelled as a LEGION NARROW PS OXIN SZ 6N RT.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Initial consignees were notified via email and overnight mail on 04/19/2022. The consignees were instructed to inspect their inventory, return affected product, and complete the response form.
Recall details
- Recall number
- Z-1099-2022
- Status
- Terminated
- Initiated by the company
- Apr 19, 2022
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 16 units
- Product code
- JWH
- FDA submissions
- K111711, K112941
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CT, NC, CO, SC, VA, FL, IL, CA and the country of Japan.
Product description
LEGION NARROW PS OXIN SZ 6N RT and JRNY II BCS FEMORAL OXIN LT SZ 5
Code information (models, lots, serial numbers)
LEGION NARROW PS OXIN SZ 6N RT: Model Number: 71421276; Lot: 21JM17389; UDI: (01)0885558234259(17)310925(10)21JM17389 JRNY II BCS FEMORAL OXIN LT SZ 5: Model Number: 74022125; Lot: 21JM18618; UDI: (01)00885556170410(17)310926(10)21JM18618
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- JOURNEY 74022125Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall27 reports
Product lines
Other devices under the same FDA submission
155 devices cleared under K111711, K112941 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74018608Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER67 reports
- JOURNEY 74027262Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall16 reports
- LEGION 71421265Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall6 reports
- JOURNEY 74027231Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER54 reports
- LEGION 71421275Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- JOURNEY 74022114Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER46 reports
- JOURNEY 74012455Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER41 reports
- JOURNEY 74027241Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER35 reports
- JOURNEY 74027242Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER30 reports
- JOURNEY 74022124Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 death27 reports
- JOURNEY 74022126Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER27 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page