SELECTSITE recallZ-1206-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated Apr 20, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm sent out a recall notification starting on 04/20/2022 via UPS 2-day delivery. Customers are asked to return all unused affected product, share the notice with all those who need to be aware within their organization or with any organization where the above-listed products have been transferred, and to complete the response form.

Recall details

Recall number
Z-1206-2022
Status
Open, Classified
Initiated by the company
Apr 20, 2022
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
1385 devices
Product code
DQY
FDA submissions
K183493
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Belgium, Denmark, France, Germany, Ireland, Italy, Netherlands, Poland, Spain, United Arab Emirates, United Kingdom.

Product description

The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a catheter dilator to facilitate deflectable catheter passage, and a guide catheter slitter to remove the deflectable catheter. The deflectable catheter is designed for placement of transvenous devices in or near the bundle of His. It features a deflecting distal section, controlled by the deflectable catheter handle, and an out-of plane curve on the distal tip. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.

Code information (models, lots, serial numbers)

Model Number: C304-HIS; GTIN: 00763000147013; Lot Numbers: 0010948864, 0010953298, 0010958150, 0010961222, 0010971148, 0011017539, 0011020263, 0011022874, 0011025202, 0011028574, 0011031177

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.