Digital Medical X-ray Imaging System, Model: uDR 596i, uDR… recallZ-1075-2022 · Shanghai United Imaging Healthcare Co., Ltd. · initiated Jul 30, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat exposure.

FDA root cause category: Software change control

What the company is doing

On 07/30/19, Customer Notification Letters started being distributed to customers. Customers were asked to contact the firm if they found an inconsistency between the protocol name and the positioning image. The following phone numbers were provided: (855) 221-1552 (toll free) or (832) 699-2799 (direct). The firm issued Field Correction Instructions to Service Engineers to modify the protocol name of Flexion and Extension on C-Spine & L-Spine with the corrected guidelines. In addition, the firm committed to permanently fix impacted systems with a software update.

Recall details

Recall number
Z-1075-2022
Status
Terminated
Initiated by the company
Jul 30, 2019
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Quantity
4
Product code
KPR
FDA submissions
K181413
Distribution
US Nationwide distribution in the state of TX.

Product description

Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h

Code information (models, lots, serial numbers)

DI/GTIN: 06971576833023, Serial Numbers: 270012, 270013,270014, 270015