Skip to content

Mammotome HydroMARK recallZ-1124-2022 · Devicor Medical Products Inc · initiated Mar 30, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Devicor Medical Products Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Small patient labels do not match the primary device label. The primary device label accurately identifies the product as 4010-05-08-T4, while the smaller patient label incorrectly identifies the product as 4010-05-08-T3.

FDA root cause category: Labeling Change Control

What the company is doing

On March 30, 2022, the firm distributed Urgent Medical Device Recall letters to their customers via email. Customers were informed of the labeling issue. Users were instructed to examine their inventory and identify any affected product. Users should contact their local sales representative for correction of any affected devices. The sales representative will be on site to remove the affected patient labels. Please do not use affected patient labels from this lot. If you have any additional questions, comments or concerns please contact the firm's Customer Support at +1 877-926-2666. You may also email [email protected].

Recall details

Recall number
Z-1124-2022
Status
Terminated
Initiated by the company
Mar 30, 2022
Recalling firm
Devicor Medical Products Inc
Quantity
90 devices
Product code
NEU
FDA submissions
K210752
Distribution
US Nationwide distribution in the states of DE, KY, NY, OH, WV.

Product description

HydroMARK Breast Biopsy Marker, Model No. 4010-05-08-T4

Code information (models, lots, serial numbers)

UDI-DI: 00841911102472 Lot F12201281D

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page