Skip to content

GETINGE CM320-SERIES recallZ-1248-2022 · Getinge Usa Sales Inc · initiated Apr 13, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection

FDA root cause category: Process control

What the company is doing

Getinge issued to U.S. consignees Medical Device Correction Letter dated April 13, 2022 via FedEx delivery. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any Getinge CM320-series Washer Disinfectors with the REF number/serial numbers listed on page 1. You can continue to use Getinge CM320-series Washer Disinfectors until the corrective action is performed; there are no precautionary actions necessary. 2. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 5) to acknowledge that you have received this notification. A Getinge service technician will review the Installation Checklist and verify that each of the steps were performed to ensure continual safe operation of your cart washer(s). This review and verification will be performed. Questions contact your Getinge representative or Technical Support please call (888) 943-8872 (options 4,2,1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). Technical Support can also be reached via email at [email protected].

Recall details

Recall number
Z-1248-2022
Status
Open, Classified
Initiated by the company
Apr 13, 2022
Recalling firm
Getinge Usa Sales Inc
Quantity
4 units
Product code
MEC
Distribution
US Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.

Product description

Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, etc. Model: CM320 WU

Code information (models, lots, serial numbers)

UDI-DI: 07340153700185 Serial Numbers: W50034570 WAA070170 WAA070618 WAA081312

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90146).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page