Atellica IM 1600 Analyzer - automated, immunoassay analyzer… recallZ-1253-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Apr 13, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lots incompatible with Test Definition (TDef) Version 1.4, potentially lead to the generation of erroneous total T3 results
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued an Urgent Medical Device Correction (UMDC) to US affected customers by FedEx beginning on 04/13/2022 and an Urgent Field Safety Notice (UFSN) was issued to all OUS. Letter states reason for recall, health risk and action to take: If your facility has multiple lots of T3 available for use and some kit lots are below lots ending in 244, follow the steps below to ensure that T3 results will be correct for all available lots until all kit lots available for use are kit lots ending in 244 or above. Ensure that all IM modules connected to an Atellica Solution have only 1 lot of T3 reagent (all IM modules have the same lot of T3 reagent packs) loaded onto every IM. For reagent kit lots below lots ending in 244: Rescan the Master Curve Card (MCC) for the T3 reagent lot that is on the Atellica IM before processing with that reagent lot, following instructions in the Atellica Online Help (SMN 11313586, section Scanning IM Master Curves and Assay Test Definitions 2D Barcodes ). Note: If the IM is rebooted while using kit lots below lots ending in 244, the MCC will need to be re-scanned before samples are processed. For reagent lots ending in 244 and above: Rescan the Master Curve Card (MCC) for the T3 reagent kit lot that is on the Atellica IM, following instructions in the Atellica Online Help (SMN 11313586, section Scanning IM Master Curves and Assay Test Definitions 2D Barcodes . Reboot all IM modules connected to Atellica Solution following instructions in the Atellica Online Help (SMN 11313586, section titled Shutting Down and Powering Off Analyzers within the System ) when changing from reagent kit lots below lots ending in 244 to a reagent pack associated with kit lots ending in 244 or above before processing with that reagent kit lot. Please review this letter with your Medical Director. Complete and return the Field Action Effectiveness Check Form attached to this letter within 30 days.
Recall details
- Recall number
- Z-1253-2022
- Status
- Open, Classified
- Initiated by the company
- Apr 13, 2022
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 2,321
- Product code
- JLW
- FDA submissions
- K151792
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Argentina, Austria, Bahamas, Bangladesh, Belgium, Brazil, Chile, China, Columbia, Croatia, Czech¿Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Latvia, Malaysia, Mexico, Netherlands, Paraguay, Peru, Philippines, Portugal, Romania, Russian¿Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United Kingdom, Vietnam.
Product description
Atellica IM 1600 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens Siemens Material Number (SMN): 11066000
Code information (models, lots, serial numbers)
UDI: 00630414002026 All lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica IM 1600 Analyzer 11066000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE7 recalls196 reports
Other devices under the same FDA submission
1 device cleared under K151792 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atellica IM 1300 Analyzer 11066001Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE6 recalls188 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90153).
- Z-1252-2022 · Open, ClassifiedAtellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page