Metalogix Slotted 90 Tubular Wrench x 10mm REF 900104 LOT… recallZ-1280-2022 · New Standard Device Inc · initiated Aug 14, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from New Standard Device Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing processing error, wrench causing stripping when torqueing nuts.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 08/19/2019, the firm sent via email an "FIELD SAFETY NOTICE" informing customers that the Slotted 90 degree Tubular Wrench was stripping when torqueing nuts in the application process. Customers were asked to: -remove the affected products from inventory and return to Metalogix Inc. at 4848 Research Drive, San Antonio, TX 78240 ATTENTION: QA Dept.

Recall details

Recall number
Z-1280-2022
Status
Open, Classified
Initiated by the company
Aug 14, 2019
Recalling firm
New Standard Device Inc
Quantity
14 devices
Product code
HXC
FDA submissions
K181630
Distribution
U.S.: AZ. CT, GA, OH, and VA O.U.S.: N/A

Product description

Metalogix Slotted 90 Tubular Wrench x 10mm REF 900104 LOT A29618A

Code information (models, lots, serial numbers)

Catalog/Part Number: 900104 UDI-DI (GTIN): 854641008112 Lot Number: A29618A

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K181630 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.