Skip to content

Artis icono floor-angiography systems developed for single… recallZ-1246-2022 · Siemens Medical Solutions USA, Inc · initiated Apr 11, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm

FDA root cause category: Device Design

What the company is doing

Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via AX027/21/S on 4/11/22. Siemens will conduct an inspection of the protective earth via Update Instruction AX026/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.

Recall details

Recall number
Z-1246-2022
Status
Open, Classified
Initiated by the company
Apr 11, 2022
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
13 units
Product code
OWB
FDA submissions
K190768, K193326
Distribution
US Nationwide distribution.

Product description

Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700

Code information (models, lots, serial numbers)

UDI-DI: 04056869149325 Serial Number: 170026 170305 170307 170309 170310 170023 170304 170025 170051 170312 170027 170303 170053

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K190768, K193326 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90184).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page