Skip to content

Clareon recallZ-1269-2022 · Alcon Laboratories Ireland, Ltd · initiated May 11, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Alcon Laboratories Ireland, Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to device (preloaded intraocular lens) mislabeling, associated with partial mixing of product during overlapping manufacturing runs, and attributed to improper manufacturing line clearance.

FDA root cause category: Under Investigation by firm

What the company is doing

On 05/11/2022, Alcon Laboratories sent an "URGENT: VOLUNTARY MEDICAL DEVICE REMOVAL" Letter via FedEx overnight informing customer that the firm has initiated a removal of one specific lot (25238769) of the Clareon IOL with AutonoMe Delivery System (CNA0T0.175). Based on the firm's investigation, there is the potential for limited number of 24.0D lenses to be included within this lot of the 17.5D Clareon IOL AutonoMe Delivery System. Customers are asked to locate and return the Alcon 17.5D Clareon IOL AutonoMe Delivery System with lot 25238769. An Alcon associate will reach out to customers directly to arrange for the return and replacement of the affected products. For questions, contact Alcon Customer Service at 1-800-862-5266 or contact their Alcon Sales Representative.

Recall details

Recall number
Z-1269-2022
Status
Terminated
Initiated by the company
May 11, 2022
Recalling firm
Alcon Laboratories Ireland, Ltd
Quantity
139 lenses (Lot 25238769) and 164 lenses (Lot 25238775)
Product code
HQL
FDA submissions
P190018
Distribution
U.S.: AL, CA, FL, GA, IA, IL, KY, LA, MI, MO, NY, OH,OK, TX, VA, and WI O.U.S.: Czech Republic, France, Germany, India, Slovakia, Spain, and Switzerland

Product description

Alcon Clareon IOL with AutonoMe Delivery System

Code information (models, lots, serial numbers)

Catalog/Model Number: CNA0T0.175 UDI Code: (01) 00380652393669 (17) 240720 (21) 25238769000 (11) 210722 (240) CNA0T0175 Lot number: 25238769  Catalog/Model Number: CNA0T0.240 UDI Code: (01) 00380652393799 (17) 240720 (21) 25238775000 (11) 210722 (240) CNA0T0240 Lot Number: 25238775

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page