Skip to content

EVIS EXERA II COLONOVIDEOSCOPE recallZ-1274-2022 · Olympus Corporation of the Americas · initiated Apr 22, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding

FDA root cause category: Nonconforming Material/Component

What the company is doing

Olympus issued letter via e-mail on April 22, 2022. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide a service loaner device in an effort to not disrupt patient care. Contact: 647-999-3203 or at [email protected]

Recall details

Recall number
Z-1274-2022
Status
Terminated
Initiated by the company
Apr 22, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
1 unit
Product code
FDF
FDA submissions
K100584
Distribution
US Nationwide distribution in the state of KY.

Product description

Colonovideoscope Model CF-H180AL

Code information (models, lots, serial numbers)

UDI-DI: 04953170339615 Serial number: 2501131

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page