Skip to content

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE recallZ-1275-2022 · Olympus Corporation of the Americas · initiated Apr 22, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications

FDA root cause category: Nonconforming Material/Component

What the company is doing

Olympus issued Urgent Medical Device Correction letter via email on 4/22/22. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide for a service loaner in an effort to not disrupt patient care. Please do not hesitate to contact @ 647-999-3203 or at [email protected] for any additional information or support concerning this matter.

Recall details

Recall number
Z-1275-2022
Status
Open, Classified
Initiated by the company
Apr 22, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
8 units
Product code
FDS
Distribution
US Nationwide distribution in the states of AL, FL, NC, NY.

Product description

EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190

Code information (models, lots, serial numbers)

UDI-DI: 04953170305276 Serial: 2269931, 2611873, 2613235, 2634251, 2634326, 2860721, 2961773 and 2962545

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates