Sensica UO System recallZ-1293-2022 · C.R. Bard Inc · initiated May 19, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from C.R. Bard Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints relating to urine output measurement accuracy.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm issued a recall notification via email and Fed Ex on 05/19/2022. The letter instructs the consignee to immediately discontinue use of the system, quarantine and store to prevent usage, share the notice with users of the product, and return the response form.
Recall details
- Recall number
- Z-1293-2022
- Status
- Open, Classified
- Initiated by the company
- May 19, 2022
- Recalling firm
- C.R. Bard Inc
- Quantity
- 457 units
- Product code
- EXS
- Distribution
- US Nationwide.
Product description
Sensica Urine Output System
Code information (models, lots, serial numbers)
Show the full code information (14k characters)
UDI: (01) 0 0801741215773; Catalog Number: SCCS1002 (Lot Numbers: BMFURMSF03, BMFURMSF07, BMFURMSF08, BMFURMSF10, BMFURMSF12, BMFURMSF13, BMFURMSF15, BMFURMSF16, BMFURMSF18, BMFURMSF19, BMFURMSF20, BMFURMSF22, BMFURMSF23, BMFURMSF24, BMFURMSF25, BMFURMSF26, BMFURMSF27, BMFURMSF28, BMFURMSF29, BMFURMSF30, BMFURMSF31, BMFURMSF32, BMFURMSF33, BMFURMSF34, BMFURMSF35, BMFURMSF36, BMFURMSF37, BMFURMSF38, BMFURMSF39, BMFURMSF40, BMFURMSF41, BMFURMSF42, BMFURMSF43, BMFURMSF44, BMFURMSF45, BMFURMSF46, BMFURMSF47, BMFURMSF48, BMFURMSF49, BMFURMSF50, BMFURMSF51, BMFURMSF52, BMFURMSF53, BMFURMSF54, BMFURMSF55, BMFURMSF56, BMFURMSF57, BMFURMSF58, BMFURMSF59, BMFURMSF60, BMFVRMSQ82, BMFVRMSQ84, BMFVRMSQ85, BMFVRMSQ86, BMFVRMSQ87, BMFVRMSQ88, BMFVRMSQ92, BMFVRMSQ94, BMFVRMSQ95, BMFVRMSQ96, BMFVRMSQ97, BMFVRMSQ98, BMFVRMSQ99, BMFVRMSR00, BMFVRMSR01, BMFVRMSR02, BMFVRMSR05, BMFVRMSR06, BMFVRMSR07, BMFVRMSR08, BMFVRMSR09, BMFVRMSR10, BMFVRMSR11, BMFVRMSR12, BMFVRMSR13, BMFVRMSR14, BMFVRMSR15, BMFVRMSR19, BMFVRMSR20, BMFVRMSR22, BMFVRMSR23, BMFVRMSR24, BMFVRMSR25, BMFVRMSR26, BMFVRMSR27, BMFVRMSR28, BMFVRMSR29, BMFVRMSR31, BMFVRMSR34, BMFVRMSR36, BMFVRMSR37, BMFVRMSR38, BMFVRMSR39, BMFVRMSR40, BMFWRMSBR6, BMFWRMSBR7, BMFWRMSBR8, BMFWRMSBR9, BMFWRMSBS0, BMFWRMSBS1, BMFWRMSBS2, BMFWRMSBS3, BMFWRMSBS4, BMFWRMSBS5, BMFWRMSBS6, BMFWRMSBS7, BMFWRMSBS8, BMFWRMSBS9, BMFWRMSBT0, BMFWRMSBT1, BMFWRMSBT2, BMFWRMSBT3, BMFWRMSBT4, BMFWRMSBT5, BMFWRMSBT6, BMFWRMSBT7, BMFWRMSBT8, BMFWRMSBT9, BMFWRMSBU0, BMFWRMSBU1, BMFWRMSBU2, BMFWRMSBU3, BMFWRMSBU4, BMFWRMSBU5, BMFWRMSBU6, BMFWRMSBU7, BMFWRMSBU8, BMFWRMSBU9, BMFWRMSBV0, BMFWRMSBV1, BMFWRMSBV2, BMFWRMSBV3, BMFWRMSBV4, BMFWRMSBV5, BMFWRMSBV6, BMFWRMSBV7, BMFWRMSBV8, BMFWRMSBV9, BMFWRMSBW0, BMFWRMSBW1, BMFWRMSJ24, BMFWRMSJ25, BMFWRMSJ26, BMFWRMSJ28, BMFWRMSJ29, BMFWRMSJ30, BMFWRMSJ31, BMFWRMSJ32, BMFWRMSJ33, BMFWRMSJ34, BMFWRMSJ37, BMFWRMSJ38, BMFWRMSJ40, BMFWRMSJ42, BMFWRMSJ43, BMFWRMSJ45, BMFWRMSJ47, BMFWRMSJ48, BMFWRMSJ49, BMFWRMSJ50, BMFWRMSJ52, BMFWRMSJ53, BMFWRMSJ54, BMFWRMSJ55, BMFWRMSJ56, BMFWRMSJ57, BMFWRMSJ58, BMFWRMSJ59, BMFWRMSJ60, BMFWRMSJ61, BMFWRMSJ62, BMFWRMSJ63, BMFWRMSJ64, BMFWRMSJ65, BMFWRMSJ66, BMFWRMSJ67, BMFWRMSJ68, BMFWRMSJ69, BMFWRMSJ70, BMFWRMSJ71, BMFWRMSJ72, BMFWRMSJ73, BMFWRMSJ74, BMFWRMSJ75, BMFWRMSJ76, BMFWRMSJ77, BMFWRMSJ78, BMFWRMSJ79, BMFWRMSJ80, BMFWRMSJ81, BMFWRMSJ82, BMFWRMSJ83, BMFXRMSBA1, BMFXRMSBA2, BMFXRMSBA3, BMFXRMSBA4, BMFXRMSBA5, BMFXRMSBA6, BMFXRMSBA8, BMFXRMSBA9, BMFXRMSBB0, BMFXRMSBB1, BMFXRMSBB2, BMFXRMSBB3, BMFXRMSBB4, BMFXRMSBB5, BMFXRMSBB6, BMFXRMSBB7, BMFXRMSBB8, BMFXRMSBB9, BMFXRMSBC0, BMFXRMSBC1, BMFXRMSBC2, BMFXRMSBC3, BMFXRMSBC4, BMFXRMSBC5, BMFXRMSBC6, BMFXRMSBC7, BMFXRMSBC8, BMFXRMSBD0, BMFXRMSJ01, BMFXRMSJ02, BMFXRMSJ03, BMFXRMSJ04, BMFXRMSJ05, BMFXRMSJ06, BMFXRMSJ07, BMFXRMSJ08, BMFXRMSJ09, BMFXRMSJ10, BMFXRMSJ11, BMFXRMSJ12, BMFXRMSJ13, BMFXRMSJ14, BMFXRMSJ15, BMFXRMSJ16, BMFXRMSJ17, BMFXRMSJ18, BMFXRMSJ19, BMFXRMSJ20, BMFXRMSJ21, BMFXRMSJ22, BMFXRMSJ23, BMFXRMSJ24, BMFXRMSJ25, BMFXRMSJ26, BMFXRMSJ27, BMFXRMSJ28, BMFXRMSJ29, BMFXRMSJ30, BMFXRMSJ31, BMFXRMSJ32, BMFXRMSJ33, BMFXRMSJ34, BMFXRMSJ35, BMFXRMSJ36, BMFXRMSJ37, BMFXRMSJ38, BMFXRMSJ39, BMFXRMSJ40, BMFXRMSJ41, BMFXRMSJ42, BMFXRMSJ43, BMFXRMSJ44, BMFXRMSJ45, BMFXRMSJ46, BMFXRMSJ47, BMFXRMSJ48, BMFXRMSJ49, BMFXRMSJ50, BMFXRMSJ51, BMFXRMSJ52, BMFXRMSJ53, BMFXRMSJ54, BMFXRMSJ55, BMFXRMSJ56, BMFXRMSJ57, BMFXRMSJ58, BMFXRMSJ59, BMFXRMSJ60, BMFYRMSBA1, BMFYRMSBA2, BMFYRMSBA3, BMFYRMSBA4, BMFYRMSBA5, BMFYRMSBA6, BMFYRMSBA7, BMFYRMSBA8, BMFYRMSBA9, BMFYRMSBB0, BMFYRMSBB1, BMFYRMSBB2, BMFYRMSBB3, BMFYRMSBB4, BMFYRMSBB5, BMFYRMSBB6, BMFYRMSBB7, BMFYRMSBB8, BMFYRMSBB9, BMFYRMSBC0, BMFYRMSBC1, BMFYRMSBC2, BMFYRMSBC3, BMFYRMSBC4, BMFYRMSBC5, BMFYRMSBC6, BMFYRMSBC7, BMFYRMSBD0, BMFYRMSBD1, BMFYRMSBD2, BMFYRMSBD3, BMFYRMSBD4, BMFYRMSBD5, BMFYRMSBD6, BMFYRMSBD7, BMFYRMSBD8, BMFYRMSBD9, BMFYRMSBE0, BMFYRMSBE1, BMFYRMSBE2, BMFYRMSBE3, BMFYRMSBE4, BMFYRMSBE5, BMFYRMSBE6, BMFYRMSBE7, BMFYRMSBE8, BMFYRMSBE9, BMFYRMSBF0, BMFYRMSBF1, BMFYRMSBF2, BMFYRMSBF3, BMFYRMSBF4, BMFYRMSBF5, BMFYRMSBF6, BMFYRMSBF7, BMFYRMSBF8, BMFYRMSBF9, BMFYRMSBG0, BMFYRMSJ01, BMFYRMSJ02, BMFYRMSJ03, BMFYRMSJ04, BMFYRMSJ05, BMFYRMSJ06, BMFYRMSJ07, BMFYRMSJ08, BMFYRMSJ09, BMFYRMSJ10, BMFYRMSJ11, BMFYRMSJ12, BMFYRMSJ13, BMFYRMSJ14, BMFYRMSJ15, BMFYRMSJ16, BMFYRMSJ17, BMFYRMSJ18, BMFYRMSJ19, BMFYRMSJ20, BMFYRMSJ21, BMFYRMSJ22, BMFYRMSJ23, BMFYRMSJ24, BMFYRMSJ25, BMFYRMSJ26, BMFYRMSJ27, BMFYRMSJ28, BMFYRMSJ29, BMFYRMSJ30, BMFYRMSJ31, BMFYRMSJ32, BMFYRMSJ33, BMFYRMSJ34, BMFYRMSJ35, BMFYRMSJ36, BMFYRMSJ37, BMFYRMSJ38, BMFYRMSJ39, BMFYRMSJ40, BMFYRMSJ41, BMFYRMSJ42, BMFYRMSJ43, BMFYRMSJ44, BMFYRMSJ45, BMFYRMSJ46, BMFYRMSJ47, BMFYRMSJ48, BMFYRMSJ49, BMFYRMSJ50, BMFYRMSJ51, BMFYRMSJ52, BMFYRMSJ53, BMFYRMSJ54, BMFYRMSJ55, BMFYRMSJ56, BMFYRMSJ57, BMFYRMSJ58, BMFYRMSJ59, BMFYRMSJ60, BMFZRMS065, BMFZRMS066, BMFZRMS067, BMFZRMS068, BMFZRMS069, BMFZRMS070, BMFZRMS071, BMFZRMS072, BMFZRMS073, BMFZRMS074, BMFZRMS075, BMFZRMS076, BMFZRMS077, BMFZRMS078, BMFZRMS079, BMFZRMS080, BMFZRMS081, BMFZRMS082, BMFZRMS083, BMFZRMS084, BMFZRMS085, BMFZRMS086, BMFZRMS087, BMFZRMS088, BMFZRMS089, BMFZRMS090, BMFZRMS091, BMFZRMS092, BMFZRMS093, BMFZRMS094, BMFZRMS095, BMFZRMS096, BMFZRMS097, BMFZRMS098, BMFZRMS099, BMFZRMS100, BMFZRMS101, BMFZRMS102, BMFZRMS104, BMFZRMS105, BMFZRMS106, BMFZRMS107, BMFZRMS108, BMFZRMS109, BMFZRMS110, BMFZRMS111, BMFZRMS112, BMFZRMS113, BMFZRMS114, BMFZRMS115, BMFZRMS116, BMFZRMS117, BMFZRMS118, BMFZRMSH47, BMFZRMSH48, BMFZRMSH49, BMFZRMSH50, BMFZRMSH51, BMFZRMSH52, BMFZRMSH53, BMFZRMSH54, BMFZRMSH55, BMFZRMSH56, BMFZRMSH57, BMFZRMSH58, BMFZRMSH59, BMFZRMSH60, BMFZRMSH61, BMFZRMSH62, BMFZRMSH63, BMFZRMSH64, BMFZRMSH65, BMFZRMSH66, BMFZRMSH67, BMFZRMSH68, BMFZRMSH69, BMFZRMSH70, BMFZRMSH71, BMFZRMSH72, BMFZRMSH73, BMFZRMSH74, BMFZRMSH75, BMFZRMSH76, BMGNRMS001, BMGNRMS002, BMGNRMS003, BMGNRMS004, BMGNRMS005, BMGNRMS006, BMGNRMS007, BMGNRMS008, BMGNRMS009, BMGNRMS010, BMGNRMS011, BMGNRMS012, BMGNRMS013, BMGNRMS014, BMGNRMS015, BMGNRMS016, BMGNRMS018, BMGNRMS019, BMGNRMS020, BMGNRMS021, BMGNRMS022, BMGNRMS023, BMGNRMS024, BMGNRMS025, BMGNRMS026, BMGNRMS027, BMGNRMS028, BMGNRMS029, BMGNRMS030, BMGNRMS058, BMGNRMS059, BMGNRMS060, BMGNRMS061, BMGNRMS062, BMGNRMS063, BMGNRMS064, BMGNRMS065, BMGNRMS066, BMGNRMS067, BMGNRMS068, BMGNRMS069, BMGNRMS070, BMGNRMS071, BMGNRMS072, BMGNRMS073, BMGNRMS074, BMGNRMS075, BMGNRMS076, BMGNRMS077, BMGNRMS078, BMGNRMS079, BMGNRMS080, BMGNRMS081, BMGNRMS082, BMGNRMS083, BMGNRMS084, BMGNRMS085, BMGNRMS086, BMGNRMS087, BMGNRMSG52, BMGNRMSG53, BMGNRMSG54, BMGNRMSG55, BMGNRMSG56, BMGNRMSG57, BMGNRMSG58, BMGNRMSG59, BMGNRMSG60, BMGNRMSG61, BMGNRMSG62, BMGNRMSG63, BMGNRMSG64, BMGNRMSG65, BMGNRMSG66, BMGNRMSG67, BMGNRMSG68, BMGNRMSG69, BMGNRMSG70, BMGNRMSG71, BMGNRMSG72, BMGNRMSG73, BMGNRMSG74, BMGNRMSG75, BMGNRMSG76, BMGNRMSG77, BMGNRMSG78, BMGNRMSG79, BMGNRMSG80, BMGNRMSG81, BMFQRMYS038, BMFQRMYS040, BMFQRMYS023, BMFQRMYS028, BMFQRMYS030, BMFQRMYS033, BMFQRMYS034, BMFQRMYS037, BMFQRMYS021, BMFQRMYS022, BMFQRMYS031, BMFQRMYS032, BMFQRMYS035, BMFQRMYS036, BMFQRMYS039, BMFPRMYS005, BMFPRMYS015, BMFPRMYS020, BMFQRMYS025, BMFQRMYS026, BMFQRMYS029, BMFPRMYS003, BMFPRMYS001, BMFPRMYS002, BMFPRMYS013, BMFPRMYS014, BMFPRMYS017, BMFQRMYS024, BMFQRMYS027, BMFQRMS246, BMFPRMYS004, BMFPRMYS007, BMFPRMYS008, BMFPRMYS009, BMFPRMYS010, BMFPRMYS011, BMFPRMYS018, BMFQRMS238, BMFQRMS240, BMFQRMS244, BMFQRMS247, BMFQRMS248, BMFQRMS250, BMFQRMS253, BMFQRMS257, BMFQRMS267, BMFQRMS268, BMFQRMS272, BMFQRMS275, BMFQRMS276, BMFQRMS279, BMFQRMS280, BMFQRMS282, BMFQRMS284, BMFQRMS285, BMFQRMS286, BMFQRMS288, BMFQRMS289, BMFQRMS293, BMFQRMS294, BMFQRMS301, BMFQRMS306, BMFQRMS317, BMFQRMS324, BMFQRMS327, BMFQRMS328, BMFQRMS330, BMFQRMS334, BMFRRMS601, BMFRRMS602, BMFRRMS633, BMFRRMS645, BMFRRMS647, BMFRRMS656, BMFRRMS659, BMFRRMST97, BMFRRMSU02, BMFRRMSU05, BMFRRMSU06, BMFRRMSU09, BMFRRMSU11, BMFRRMSU12, BMFRRMSU19, BMFRRMSU24, BMFRRMSU28, BMFRRMSU34, BMFRRMSU37, BMFSRMSH84, BMFSRMSH88, BMFSRMSJ05, BMFRRMSU23, BMFPRMYS006, BMFPRMYS012, BMFQRMS241, BMFQRMS245, BMFQRMS255, BMFQRMS259, BMFQRMS270, BMFQRMS281, BMFQRMS283, BMFQRMS287, BMFQRMS290, BMFQRMS291, BMFQRMS292, BMFQRMS295, BMFQRMS296, BMFQRMS299, BMFQRMS300, BMFQRMS307, BMFQRMS308, BMFQRMS315, BMFQRMS320, BMFQRMS322, BMFQRMS325, BMFQRMS326, BMFQRMS329, BMFQRMS331, BMFQRMS332, BMFQRMS335, BMFRRMS603, BMFRRMS610, BMFRRMS621, BMFRRMS631, BMFRRMS646, BMFRRMS649, BMFRRMS657, BMFRRMS658, BMFRRMST91, BMFRRMST92, BMFRRMST96, BMFRRMST99, BMFRRMSU03, BMFRRMSU04, BMFRRMSU08, BMFRRMSU10, BMFRRMSU13, BMFRRMSU14, BMFRRMSU16, BMFRRMSU17, BMFRRMSU20, BMFSRMSH89, BMFSRMSJ35, BMFSRMSJ39, BMFRRMSU21, BMFRRMSU31, BMFRRMSU32, BMFRRMSU35, BMFRRMSU36, BMFRRMSU38, BMFRRMSU39, BMFRRMSU42, BMFRRMSU45, BMFSRMSH86, BMFQRMS231, BMFQRMS232, BMFQRMS235, BMFQRMS236, BMFQRMS239, BMFQRMS249, BMFQRMS251, BMFQRMS263, BMFQRMS264, BMFQRMS265, BMFQRMS273, BMFQRMS304, BMFQRMS310, BMFQRMS313, BMFQRMS314, BMFQRMS318, BMFQRMS321, BMFRRMS613, BMFRRMS617, BMFRRMS619, BMFRRMS622, BMFRRMS625, BMFRRMS632, BMFRRMS650, BMFRRMS651, BMFRRMS653, BMFRRMS655, BMFRRMST94, BMFRRMST98, BMFRRMSU01, BMFRRMSU15, BMFRRMSU25, BMFSRMSH91, BMFSRMSH92, BMFSRMSJ04, BMFSRMSJ08, BMFSRMSJ12, BMFSRMSJ16, BMFSRMSJ24, BMFSRMSJ40, BMFRRMSU27, BMFRRMSU29, BMFRRMSU33, BMFRRMSU40, BMFRRMSU44, BMFRRMSU47, BMFRRMSU48, BMFRRMSU50, BMFSRMSH72, BMFSRMSH76, BMFQRMS233, BMFQRMS234, BMFQRMS237, BMFQRMS242, BMFQRMS243, BMFQRMS252, BMFQRMS254, BMFQRMS256, BMFQRMS258, BMFQRMS260, BMFQRMS261, BMFQRMS266, BMFQRMS269, BMFQRMS271, BMFQRMS274, BMFQRMS297, BMFQRMS298, BMFQRMS302, BMFQRMS309, BMFQRMS319, BMFQRMS323, BMFQRMS333, BMFRRMS606, BMFRRMS607, BMFRRMS608, BMFRRMS614, BMFRRMS615, BMFRRMS618, BMFRRMS620, BMFRRMS623, BMFRRMS624, BMFRRMS652, BMFRRMS654, BMFRRMST93, BMFRRMST95, BMFRRMSU00, BMFRRMSU07, BMFRRMSU18, BMFRRMSU22, BMFRRMSU26, BMFRRMSU30, BMFRRMSU41, BMFRRMSU43, BMFRRMSU46, BMFRRMSU49, BMFRRMSU51, BMFRRMSU52, BMFSRMSH73, BMFSRMSH74, BMFSRMSH77, BMFSRMSH85, BMFSRMSH87, BMFSRMSH90, BMFSRMSH96, BMFSRMSJ41, BMFSRMSJ45, BMFSRMSJ46, BMFQRMS277, BMFQRMS278, BMFQRMS303, BMFRRMS604, BMFRRMS626, BMFRRMS629, BMFRRMS635, BMFRRMS636, BMFRRMS639, BMFRRMS641, BMFRRMS642, BMFSRMSJ11, BMFQRMS262, BMFQRMS305, BMFQRMS311, BMFQRMS312, BMFQRMS316, BMFRRMS605, BMFRRMS609, BMFRRMS611, BMFRRMS612, BMFRRMS616, BMFRRMS627, BMFRRMS628, BMFRRMS630, BMFRRMS634, BMFRRMS637, BMFRRMS638, BMFRRMS640, BMFRRMS643, BMFRRMS644, BMFRRMS648, BMFSRMSJ03, BMFSRMSJ10, BMFSRMSJ29, BMFSRMSH78, BMFSRMSH81, BMFSRMSH82, BMFSRMSJ00, BMFSRMSJ14, BMFSRMSH79, BMFSRMSH80, BMFSRMSH83, BMFSRMSH93, BMFSRMSJ15, BMFSRMSJ13, BMFSRMSJ27, BMFSRMSJ33, BMFSRMSJ34, BMFSRMSH75, BMFSRMSJ02, BMFSRMSJ06, BMFSRMSJ09, BMFSRMSJ19, BMFSRMSJ23, BMFSRMSJ28, BMFSRMSJ30, BMFSRMSJ07, BMFSRMSJ17, BMFSRMSJ21, BMFSRMSJ22, BMFSRMSJ25, BMFSRMSJ26, BMFSRMSJ32, BMFSRMSJ31, BMFSRMSJ18, BMFSRMSJ20, BMFSRMSJ43, BMFTRMS121, BMFSRMSJ38, BMFSRMSJ44, BMFSRMSJ37, BMFTRMS108, BMFTRMS114, BMFTRMS142, BMFSRMSJ36, BMFSRMSJ42, BMFTRMS106, BMFTRMS109, BMFTRMS138, BMFSRMSH94, BMFSRMSH99, BMFSRMSH97, BMFTRMS122, BMFTRMS135, BMFTRMS136, BMFVRMSQ68, BMFSRMSH95, BMFSRMSH98, BMFSRMSJ01, BMFTRMS120, BMFTRMS123, BMFTRMS137, BMFVRMSQ62, BMFVRMSQ66, BMFTRMS140, BMFVRMSQ63, BMFVRMSQ64, BMFVRMSQ77, BMFVRMSQ78, BMFVRMSQ65, BMFVRMSQ72, BMFVRMSQ75, BMFVRMSQ76, BMFVRMSQ80, BMFTRMS102, BMFTRMS105, BMFTRMS111, BMFTRMS112, BMFTRMS115, BMFTRMS116, BMFTRMS119, BMFTRMS124, BMFTRMS127, BMFTRMS128, BMFTRMS090, BMFTRMS093, BMFTRMS094, BMFTRMS098, BMFTRMS101, BMFVRMSQ60, BMFVRMSQ67, BMFTRMS130, BMFTRMS133, BMFTRMS134, BMFTRMS143, BMFTRMS144, BMFTRMS147, BMFTRMS148, BMFVRMSQ55, BMFVRMSQ56, BMFVRMSQ59, BMFTRMS089, BMFTRMS091, BMFTRMS092, BMFTRMS095, BMFTRMS096, BMFTRMS099, BMFTRMS100, BMFTRMS103, BMFTRMS104, BMFTRMS107, BMFTRMS110, BMFTRMS113, BMFTRMS117, BMFTRMS118, BMFTRMS125, BMFTRMS126, BMFTRMS129, BMFTRMS131, BMFTRMS132, BMFTRMS139, BMFTRMS141, BMFTRMS145, BMFTRMS146, BMFVRMSQ53, BMFVRMSQ57, BMFVRMSQ61, BMFVRMSQ69, BMFVRMSQ71, BMFVRMSQ91, BMFWRMSJ35, BMFVRMSQ51, BMFVRMSQ52, BMFVRMSQ81, BMFVRMSQ89, BMFVRMSQ54, BMFVRMSQ79, BMFVRMSQ83, BMFVRMSQ90, BMFWRMSJ39, BMFWRMSJ41, BMFWRMSJ44, BMFVRMSQ58, BMFWRMSJ27, BMFVRMSQ70, BMFVRMSQ73, BMFVRMSQ74, BMFVRMSR17, BMFVRMSR18, BMFVRMSR30, BMFVRMSR33, BMFVRMSR21, BMFVRMSR32, BMFVRMSR16, BMFURMSF17, BMFVRMSR35, BMFURMSF04, BMFURMSF14, BMFURMSF21, BMFURMSF02, BMFURMSF11, BMFURMSF01, BMFURMSF05, BMFWRMSJ36, BMFWRMSJ51, BMFURMSF06, BMFURMSF09, BMFVRMSR03, BMFVRMSR04 ) ; SCCS1002E (Lot Numbers: BMFQRMYS001, BMFQRMYS002, BMFQRMYS003, BMFQRMYS004, BMFQRMYS005, BMFQRMYS006, BMFQRMYS007, BMFQRMYS008, BMFQRMYS009, BMFQRMYS010, BMFQRMYS011, BMFQRMYS012, BMFQRMYS013, BMFQRMYS014, BMFQRMYS015, BMFQRMYS016, BMFQRMYS017, BMFQRMYS019, BMFQRMYS041, BMFQRMYS042, BMFQRMYS043, BMFQRMYS044, BMFQRMYS045, BMFQRMYS046, BMFQRMYS047, BMFQRMYS048, BMFQRMYS049, BMFQRMYS050, BMFQRMYS051, BMFQRMYS052, BMFQRMYS053, BMFQRMYS054, BMFQRMYS055, BMFQRMYS056, BMFQRMYS057, BMFQRMYS058, BMFQRMYS059, BMFQRMYS060, BMFQRMYS061, BMFQRMYS062, BMFQRMYS063, BMFQRMYS064, BMFQRMYS065, BMFQRMYS066, BMFQRMYS067, BMFQRMYS068, BMFQRMYS069, BMFQRMYS070, BMFQRMYS071, BMFQRMYS072, BMFQRMYS073, BMFQRMYS074, BMFQRMYS075, BMFQRMYS076, BMFQRMYS077, BMFQRMYS078, BMFQRMYS079, BMFQRMYS080, BMFQRMYS081, BMFQRMYS082, BMFQRMYS083, BMFQRMYS084, BMFQRMYS085, BMFQRMYS086, BMFQRMYS087, BMFQRMYS088, BMFQRMYS089, BMFQRMYS090, BMFQRMYS091, BMFQRMYS092, BMFQRMYS093, BMFQRMYS094, BMFQRMYS095, BMFQRMYS096, BMFQRMYS097, BMFQRMYS098, BMFQRMYS099, BMFQRMYS100, BMFQRMYS101, BMFQRMYS102, BMFQRMYS103, BMFQRMYS104, BMFQRMYS105, BMFQRMYS106, BMFQRMYS107, BMFQRMYS108, BMFQRMYS109, BMFQRMYS110, BMFQRMYS111, BMFQRMYS112, BMFQRMYS113, BMFQRMYS114, BMFQRMYS115, BMFQRMYS116, BMFQRMYS117, BMFQRMYS118, BMFQRMYS119, BMFQRMYS120, BMFQRMYS121, BMFQRMYS122, BMFQRMYS123, BMFQRMYS124, BMFQRMYS125, BMFQRMYS126, BMFQRMYS127, BMFQRMYS128, BMFQRMYS129, BMFQRMYS130, BMFQRMYS131, BMFQRMYS132, BMFQRMYS133, BMFQRMYS134, BMFQRMYS135, BMFQRMYS136, BMFQRMYS137, BMFQRMYS138, BMFQRMYS139, BMFQRMYS140, BMFQRMYS141, BMFQRMYS142, BMFQRMYS143, BMFQRMYS144, BMFQRMYS145, BMFQRMYS146, BMFQRMYS147, BMFQRMYS148, BMFQRMYS149, BMFQRMYS150, BMFQRMYS151, BMFQRMYS152, BMFQRMYS153, BMFQRMYS154, BMFQRMYS155, BMFQRMYS156, BMFQRMYS157, BMFQRMYS158, BMFQRMYS159, BMFQRMYS161, BMFQRMYS162, BMFQRMYS163, BMFQRMYS164, BMFQRMYS165, BMFQRMYS166, BMFQRMYS167, BMFQRMYS168, BMFQRMYS169, BMFQRMYS170)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Sensica UO System SCCS1002C. R. Bard, Inc. · URINOMETER, ELECTRICAL2 recalls289 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page