Why it was recalled

Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications

FDA root cause category: Process control

What the company is doing

On May 25, 2022, Philips issued a "Urgent Medical Device Correction" notification. Phillips asked customers to take the following actions: 1. Philips recommends reviewing the enclosed compliant Instructions for Use, in Appendix A, before proceeding as normal. 2. As per standard practice, precaution should always be used when advancing or removing a catheter in complex vessel anatomies, such as not forcing a catheter into a narrow vessel or a tight stenosis. Vessel calcification, tortuosity and pre-existing untreated vessel spasm are key components to any complex vascular anatomy. Additionally, if the initial guidewire faces resistance or requires further manipulations, please use this as an indicator of complex patient anatomy. 3. Philips is providing this Urgent Medical Device Correction letter to affected customers and recommends forwarding this notification to all staff who need to be aware within your organization or to any organization where the potentially affected devices may have been transferred. 4. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form within 30 days upon receipt of this notice to Email: [email protected]

Recall details

Recall number
Z-1301-2022
Status
Open, Classified
Initiated by the company
May 13, 2022
Recalling firm
Volcano Corp
Quantity
14.031 units
Product code
OBJ
FDA submissions
K121273
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, FR, PR, IT, DE, CA, NZ, PL, BR, ZA, NL.

Product description

Volcano Visions PV. 035 Digital IVUS Catheter REF 88901, PN: 300005384002 Rx Only, Sterile EO, CE 2797

Code information (models, lots, serial numbers)

All Lots containing IFU Part Number 300004856772/A Revision Date: 05/2021 , ; UDI/DI: 00845225002329

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90252).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page