Skip to content

MassCheck Amino Acid, Acylcarnitines Dried Blood Spot… recallZ-1337-2022 · Chromsystems Instruments & Chemicals GmbH · initiated May 19, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Chromsystems Instruments & Chemicals GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products.

FDA root cause category: Labeling Change Control

What the company is doing

The firm issued a recall notification on 05/19/2022 via email. The notification mentions if the consignee has used the affected product, check whether they have determined glycine, replace the old package insert with the updated package insert with release date 18.05.2022, forward a copy of the letter to another laboratory if the affected product was given to them, and submit the response form by 01.06.2022. If users determined glycine, they are asked to clarify with the attending physician or laboratory director, if necessary, whether a back-check of the values for glycine is required.

Recall details

Recall number
Z-1337-2022
Status
Terminated
Initiated by the company
May 19, 2022
Recalling firm
Chromsystems Instruments & Chemicals GmbH
Quantity
1457 units
Product code
JJY
Distribution
US: NY, OH, CT.

Product description

MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level II

Code information (models, lots, serial numbers)

Model: 0193; Lot Number: 2821; UDI/DI: 04250317502865

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 90266).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page