Artis Icono, Interventional Fluoroscopic X-Ray System… recallZ-1342-2022 · Siemens Medical Solutions USA, Inc · initiated May 20, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During system tests, an increased wearing of the Image acquisition system fans has been observed. This could lead to an electrical failure which may cause malfunction of the image system during regular system operation.

FDA root cause category: Process control

What the company is doing

On May 20, 2022, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were informed that in the event this issue occurs, the system will switch to Bypass fluoroscopy mode. In Bypass fluoroscopy mode a limited imaging functionality (non-subtracted, continuous fluoroscopy with reduced power and without acquisition and storage of images) would remain available. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. Customer Instructions: In the event you recognize a high squeaking scratching loud noise out of one of the system cabinets, please contact your service partner for immediate action. It is recommended to avoid performing any critical interventions on the system. Due to the possible risk of electrical shock to the technician, please ensure that your service partner (if not Siemens Healthineers) first contacts Siemens Healthineers' service organization before starting any corrective actions. The firm will replace the affected parts. The service organization will contact customers to arrange a date to perform the corrective action. To schedule an earlier appointment, contact the firm at 1-800-888-7436.

Recall details

Recall number
Z-1342-2022
Status
Open, Classified
Initiated by the company
May 20, 2022
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
195 worldwide, 88 US
Product code
JAK
FDA submissions
K193326
Distribution
Domestic distribution to AL, AR, AZ, CA, CO, CT, DE, FL, ID, IL, IN, KY, LA, MA, MD, MI, MO, MT, NC, NE, NJ, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, WI and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Japan, Kazakhstan, Netherlands, New Zealand, Norway, Oman, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.

Product description

Artis Icono, Interventional Fluoroscopic X-Ray System, Model No. 11327600

Code information (models, lots, serial numbers)

UDI-DI: 04056869063317 Serial Numbers: 180062 180054 180019 180065 180056 180057 180041 180317 180034 180072 180349 180367 180382 180060 180399 180337 180070 180048 180083 180369 180370 180371 180372 180345 180315 180343 180080 180301 180305 180067 180330 180040 180379 180386 180360 180333 180029 180352 180336 180407 180400 180397 180316 180068 180363 180084 180334 180018 180350 180376 180354 180046 180047 180335 180085 180035 180339 180357 180358 180361 180362 180081 180331 180381 180021 180342 180321 180309 180356 180378 180393 180368 180341 180323 180329 180043 180049 180377 180077 180086 180053 180055 180015 180014 180066 180082 180037 180373