Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for… recallZ-1367-2022 · Ortho-Clinical Diagnostics, Inc. · initiated May 26, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).
FDA root cause category: Incorrect or no expiration date
What the company is doing
Ortho Clinical Diagnostics issued Important Correction Notification letter (CL2022-135) on May 26, 2022. Letter states reason for recall, health risk and action to take: When using VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack, do not store on-analyzer for longer than 6 weeks. " Until a lot update occurs, OAS for VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack will temporarily require manual monitoring and removal of reagent packs from analyzer no longer than 6 weeks after packs have been opened and loaded onto the analyzer. " Complete the enclosed Confirmation of Receipt form no later than June 9, 2022. " Post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System in your laboratory until further notice. " Please forward this notification if VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311
Recall details
- Recall number
- Z-1367-2022
- Status
- Open, Classified
- Initiated by the company
- May 26, 2022
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 20858
- Product code
- QKO
- Distribution
- Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.
Product description
Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems. Product Code: 6199960
Code information (models, lots, serial numbers)
UDI-DI :10758750034413. All expired, in-date and future released lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6199960Ortho Clinical Diagnostics · Reagent, coronavirus serological1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.