epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic… recallZ-1372-2022 · Siemens Healthcare Diagnostics Inc · initiated Jun 1, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens has issued an Urgent Medical Device Correction letter POC 22-010.A.US via FedEx beginning on June 1, 2022. An Urgent Field Safety Notice (UFSN) was issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: Please dispose any inventory of epoc BGEM test card lot 11-22005-50 currently in your possession in accordance with local and state disposal requirements. You may request free of charge replacement product from your local Siemens or distributor office. Please review your inventory of this product and assess your laboratory s replacement needs. Please use an epoc BGEM test card from another unaffected lot. Review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check and Product Replacement Form attached to this letter. Please retain this letter with your laboratory records and forward it to those who may have received this product. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.

Recall details

Recall number
Z-1372-2022
Status
Open, Classified
Initiated by the company
Jun 1, 2022
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
263 Boxes
Product code
CHL
FDA submissions
K192240
Distribution
AL CA CO FL IA ID KS NM OK TX

Product description

epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood Product Siemens SMN#/Model #: 10736515

Code information (models, lots, serial numbers)

UDI-DI: 00809708121860 Lot Number: 11-22005-50  Exp. Date: 22-June-2022

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K192240 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.