TRITON recallZ-1402-2022 · Stryker Instruments Div. of Stryker Corporation · initiated Jan 22, 2020

Class II Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Instruments Div. of Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.

FDA root cause category: Process control

What the company is doing

Customers were notified via email dated 1/22/20. We request the following actions from you: 1. Please confirm receipt of this communication by emailing myself and/or your Gauss representative. 2. Please identify any of these miscolored Canister Scanning Labels in your possession: " Find the the miscolored labels by locating any package of insert + scanning label with a date of manufacture prior to to 2019-12-24. See an example below: 3. Once identified, please: " Quarantine this product and indicate this has been completed by written email response to this communication Your Gauss representative will be in contact with you this week to help resolve this issue. Replacement product will be provided immediately. Thank you for your attention to this matter. Please feel free to reach out directly to me if you have any questions.

Recall details

Recall number
Z-1402-2022
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 22, 2020
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Quantity
16,700 units
Product code
PBZ
FDA submissions
K142801
Distribution
US Nationwide

Product description

Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.

Code information (models, lots, serial numbers)

UDI-DI GTIN: 10859506006101, 10859506006125 All Canister Scanning Labels with a date of manufacture prior to to 2019-12-24