Skip to content

nordicBrainEX, software versions 2.21 through 2.3.10 recallZ-1462-2022 · NordicNeuroLab AS · initiated Jun 16, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from NordicNeuroLab AS or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.

FDA root cause category: Software design

What the company is doing

The firm sent a notice regarding a Critical Bug in nordicBrainEx version 2.2.1 to 2.3.10 on 06/16/2022 by email. The notice explained the problem and advised users to use extra caution when viewing the exported results. A software revision will be released to correct the issue.

Recall details

Recall number
Z-1462-2022
Status
Terminated
Initiated by the company
Jun 16, 2022
Recalling firm
NordicNeuroLab AS
Quantity
243 units
Product code
LLZ
FDA submissions
K133910
Distribution
US Nationwide distribution.

Product description

nordicBrainEX, software versions 2.21 through 2.3.10

Code information (models, lots, serial numbers)

UDI 07090042059015, software versions 2.21 through 2.3.10

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page