Endo-Model SL Knee Prosthesis System recallZ-1379-2022 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated May 24, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On May 24, 2022, the firm notified affected customers via Urgent Field Safety Notice email. Customers were provided a modified surgical technique providing information for the cementing technique for Tilastan proximal spacers to minimize the risk. No product return is required.
Recall details
- Recall number
- Z-1379-2022
- Status
- Open, Classified
- Initiated by the company
- May 24, 2022
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Quantity
- 3785
- Product code
- JWH
- FDA submissions
- K212742
- Distribution
- Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
Product description
Tibial Component. orthopedic prosthesis. Model Nos: Small, W 16-2817/02 Medium, W 16-2817/05 Large, W 16-2817/07
Code information (models, lots, serial numbers)
UDI-DI: 04026575359202 Small, W; 04026575359219 Medium, W; 04026575359226 Large, W; Serial/Lot Numbers: 201002/1683 211025/1543 210202/0275 210302/0308 210314/1872 201002/1687
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Endo-Model SL Knee Prosthesis System 16-2817/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/07WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
Product lines
Other devices under the same FDA submission
31 devices cleared under K212742 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Endo-Model Modular Knee Prosthesis System 15-3816/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3817/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls5 reports
- Endo-Model Modular Knee Prosthesis System 15-3818/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls2 reports
- Endo-Model Modular Knee Prosthesis System 15-3815/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model Modular Knee Prosthesis System 15-3817/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Unicondylar Sled Prosthesis 15-2220/52WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Endo-Model Modular Knee Prosthesis System 15-3818/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer1 recall1 report
- Endo-Model Modular Knee Prosthesis System 15-3815/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2220/40WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2220/46WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2220/60WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90477).
- Z-1377-2022 · Open, ClassifiedEndo Model Modular Rotational Tibia - Small (Model No #15-2814/02), Medium (Model No #15-2814/03), and Large (Model No #15-2814/04). orthopedic prosthesis.
- Z-1378-2022 · Open, ClassifiedEndo Model Modular Femur and Tibia, PorEx. orthopedic prosthesis. Model Nos.: X-Small, Right 15-3815/11, Small, Right 15-3816/11, Small, Left 15-3816/12, …
- Z-1380-2022 · Open, ClassifiedTibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page