Skip to content

ENNOVATE recallZ-1428-2022 · Aesculap Implant Systems LLC · initiated Jul 13, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aesculap Implant Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system.

FDA root cause category: Component design/selection

What the company is doing

An Urgent Field Safety Notification was issued July 13, 2020 sent USPS certified mail to distributors. The notification identified the product as Ennovate Removal Key (SZ380R) and addressed correct usage of said device, referencing section K of the Surgical Technique. Distributors are to identify and notify customers if further distributed. Distributors and customers are to review the notice, notify end users, and return the "Medical Device Field Safety Notification Form" to Aesculap via email at [email protected].

Recall details

Recall number
Z-1428-2022
Status
Terminated
Initiated by the company
Jul 13, 2020
Recalling firm
Aesculap Implant Systems LLC
Quantity
16
Product code
LXH
Distribution
United States Nationwide distribution in the states of CA, CO, IL, MI, OH & TX.

Product description

ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R

Code information (models, lots, serial numbers)

UDI-DI: 04046964719622

Devices named in this recall