Skip to content

Sensis Vibe Combo with software version VD12A. A… recallZ-1426-2022 · Siemens Medical Solutions USA, Inc · initiated Jun 15, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.

FDA root cause category: Device Design

What the company is doing

Siemens issued a Customer Safety Advisory Notice (AX016/22/S) dated 6/15/22 distributed to customers. Letter states reason for recall, health risk and action to take. The BIOS configuration of the ComboBox in the affected systems will be updated to correct the issue. This corrective action will be distributed via Update Instruction AX015/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein

Recall details

Recall number
Z-1426-2022
Status
Open, Classified
Initiated by the company
Jun 15, 2022
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
13 units
Product code
DQK
FDA submissions
K150493
Distribution
US Nationwide Distribution AZ FL KY MI OH PR TX

Product description

Sensis Vibe Combo with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology Model: 11007642

Code information (models, lots, serial numbers)

UDI-DI: 04056869010205  Serial Numbers:  123021, 123055, 123056, 123058, 122042, 123004, 122006, 122026, 122029, 123047, 123053, 122011, 121014.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K150493 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90504).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page