TruClear recallZ-1467-2022 · Covidien Llc · initiated Jun 7, 2022

Class II Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.

FDA root cause category: Device Design

What the company is doing

Medtronic issued Urgent Medical Device Recall letter on 6/7/22 (Field Action #: FA1259) via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take: "Identify and quarantine all affected product. The enclosed Customer Notification DetailReport includes affected product for your facility. " Return all affected product in your inventory to Medtronic. a. Product purchased directly from Medtronic: [email protected] b. Product purchased through distributor: [email protected] " Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Complete the enclosed Customer Confirmation Form and email it to [email protected] " If you have questions or concerns regarding this recall, please contact Medtronic CustomerService at 800-854-3570.

Recall details

Recall number
Z-1467-2022
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 7, 2022
Recalling firm
Covidien Llc
Quantity
33 units
Product code
HIH
FDA submissions
K031787, K132015
Distribution
AZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom Ireland Netherlands Reunion Spain

Product description

TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808

Code information (models, lots, serial numbers)

UDI-Device Identifier (GTIN/UPN) 10884521749931 Serial Number: YF1120, YF1126, YF1127, YF1128, YF1129, YF1130, YF1131, YF1133, YF1134, YF1135, YF1137, YF1138, YF1139, YF1140, YF1141, YF1142, YF1150, YF1152, YF1154, YF1156, YF1157, YF1158, YF1161, YF1163, YF1166, YF1167, YF1168, YF1169, YF1172, YF1175, YF1176, YF1177, YF1182

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K031787, K132015 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page