Skip to content

ENNOVATE recallZ-1429-2022 · Aesculap Implant Systems LLC · initiated Nov 18, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aesculap Implant Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system. Redesign of removal key.

FDA root cause category: Component design/selection

What the company is doing

An Urgent Field Safety Notification was issued November 18, 2020 sent USPS certified mail to distributors regarding removal of Ennovate Removal Key Short (SZ380) from the field. The affected Ennovate MIS Removal Key Short (SZ380R) in the field is to be exchanged with an improved design version under the same article number manufactured after June 01, 2020. Distributors are to identify and notify customers of this action. Distributors and customers are to identify the affected version in inventory and return it to Aesculap and submit the Acknowledgment Form via email at [email protected].

Recall details

Recall number
Z-1429-2022
Status
Terminated
Initiated by the company
Nov 18, 2020
Recalling firm
Aesculap Implant Systems LLC
Quantity
17
Product code
LXH
Distribution
United States Nationwide distribution in the states of CA, CO, MI, NH, OH, SC & TX.

Product description

ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R

Code information (models, lots, serial numbers)

UDI-DI: 04046964719622

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates