Skip to content

AESCULAP S4 Cervical Cross Connector L 42-58mm, S4C CROSS… recallZ-1448-2022 · Aesculap Implant Systems LLC · initiated Sep 8, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aesculap Implant Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.

FDA root cause category: Device Design

What the company is doing

An Urgent Medical Device Removal Notification issued September 08, 2020 sent via USPS certified mail. The notice identifies the affected part numbers S4C CROSS CONNECTOR SW112T, SW113T, SW114T, SW115T, SW116T and SW117T and asks distributors and customers to identify existing inventory, quarantine, and return affected inventory. If further distributed, identify and notify customers. The notice includes an acknowledgement form to be completed and returned to Aesculap Inc. via email at [email protected]. Questions are to be directed to Lindsay Chromiak, Manager, QA Compliance and Vigilance (610) 984-9072.

Recall details

Recall number
Z-1448-2022
Status
Terminated
Initiated by the company
Sep 8, 2020
Recalling firm
Aesculap Implant Systems LLC
Quantity
27
Product code
NKB
FDA submissions
K062012
Distribution
United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.

Product description

AESCULAP S4 Cervical Cross Connector L 42-58mm, S4C CROSS CONNECTOR VARIABLE 42-58MM, product code SW117T. Used in spinal fusion surgery.

Code information (models, lots, serial numbers)

UDI-DI: 04038653362349

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 90502).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page