Skip to content

Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023… recallZ-1478-2022 · Stryker Instruments · initiated Jun 24, 2022

Class II Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Instruments or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.

FDA root cause category: Process change control

What the company is doing

On June 24, 2022, the firm notified customers via Urgent Medical Device Recall Letters. Customers were instructed to locate and quarantine any affected products found in their inventory. Upon receiving the customer's completed response form, the firm will provide a shipping label to return any recalled product. Upon receipt of the recalled product, Stryker will contact the customer to arrange for replacement product.

Recall details

Recall number
Z-1478-2022
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 24, 2022
Recalling firm
Stryker Instruments
Quantity
543
Product code
HBE
FDA submissions
K143320
Distribution
Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.

Product description

Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates