Skip to content

SteriSpine PS kit of 2 Multi-Axial Fenestrated Screws… recallZ-1572-2022 · Safe Orthopaedics LLC · initiated Jun 3, 2022

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Safe Orthopaedics LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The type of screw printed on the traceability record for the patient label is incorrect, but the type of screw printed on the other labeling is correct.

FDA root cause category: Labeling Change Control

What the company is doing

The recall letter was issued on 6/3/2022 via email to the firm's one U.S. customer. The email included a dealer recall letter, dealer response form, and a template letter for notification of any hospitals who received the product from the dealer. The letter requested the products be quarantined and the dealer should contact the hospitals who received the product to notify them of the incorrect patient labels.

Recall details

Recall number
Z-1572-2022
Status
Completed
Initiated by the company
Jun 3, 2022
Recalling firm
Safe Orthopaedics LLC
Quantity
8 devices
Product code
NKB
FDA submissions
K201648
Distribution
Distribution was made to MD. There was no government/military distribution.

Product description

SteriSpine PS kit of 2 Multi-Axial Fenestrated Screws, model numbers/REF numbers KITM640, KITM645, and KITM740.

Code information (models, lots, serial numbers)

KITM640: Lot numbers BF0A07 and 15BF001/R10; UDI 03760219910053; KITM645: Lot number BG0A02; UDI 03760219910060; and KITM740: Lot number BL0A01; UDI 03760219910091.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page