ATTUNE recallZ-1419-2022 · DePuy Orthopaedics, Inc. · initiated Jun 29, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
FDA root cause category: Process control
What the company is doing
DePuy Synthes issued Urgent Medical Device Correction Letter on June 29 2022. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately for the subject lots and quarantine the products. 2. To minimize disruption, replacement of the spring will be completed by a Field Service Representative. You will be contacted directly by a member of the Field Service Team. 3. Familiarize yourself with the content of this letter and forward this notice to any personnel in your facility who need to be informed. 4. Complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to [email protected] within three (3) business days of receipt of this medical device correction letter. Please include FA 2126657 - Attune Measured Sizing & Rotation Guide in the email subject line. 5. Complete the BRF even if you do not have any of the subject product in your current inventory. 6. If any of the subject product has been forwarded to another facility, contact that facility, and provide a copy of this notice to the relevant personnel. 7. Post a copy of this notice in a visible area for awareness of this correction. Contact your local DePuy Synthes Sales Consultant. For Medical Information Requests, please visit our website: https://www.jnjmedicaldevices.com/mir
Recall details
- Recall number
- Z-1419-2022
- Status
- Open, Classified
- Initiated by the company
- Jun 29, 2022
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 872 units
- Product code
- HWT
- Distribution
- Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
Product description
ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525
Code information (models, lots, serial numbers)
GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ATTUNE 2544-00-525Depuy (Ireland) · TEMPLATE1 recall187 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page