Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use… recallZ-1498-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 28, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Falsely Elevated Atellica CH Microalbumin_2 (ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued Urgent Medical Device Correction (UMDC) Letter by FedEx beginning on 06/28/2022. Letter states reason for recall, health risk and action to take: For customers operating with Atellica Software v.1.25.2 and lower, the resolution of this issue will be implemented in Atellica Software v1.25.3 which will be available soon. In the interim, please follow the instructions in the Additional Information section. Customers who are operating with Atellica Software v1.26 will receive further information when a software update to resolve the issue is available. For laboratories operating with Atellica Software v1.25.2 and below and Atellica Software v1.26 follow the workaround instructions in the Additional Information section until a future version of software is available. If your laboratory has multiple Atellica CH 930 Analyzers, Siemens recommends testing the Atellica CH ALB_2 assay on a separate analyzer from the Iron_2 assay. " If you choose not to separate the assays as indicated above, batch testing of Atellica CH ALB_2 may be considered. " If Iron_2 and ALB_2 will be processed on the same Atellica CH analyzer, an RPC2 wash mitigation must be initiated after processing Iron_2 and prior to processing ALB_2. Note: Any of the following will initiate the RPC2 wash: " After the Atellica CH 930 Analyzer has been in standby for 12 minutes. " After completion of any Open Channel assay. " Restarting the Atellica CH 930 Analyzer. Refer to the Atellica Solution Online Help for instructions on system restart Please review this letter with your Medical Director. Perform the instructions provided in Additional Information. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support
Recall details
- Recall number
- Z-1498-2022
- Status
- Open, Classified
- Initiated by the company
- Jun 28, 2022
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1614 units US ; 7,979 units OUS
- Product code
- JIY
- FDA submissions
- K991576
- Distribution
- Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE, BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ, IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT, QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.
Product description
Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.
Code information (models, lots, serial numbers)
Unique Device Identification (UDI): 00630414596402 All lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH Iron_2 11097601Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall1 report
Other devices under the same FDA submission
6 devices cleared under K991576 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA® Chemistry IRON_2 Iron 2 Reagents 10377510Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall1 report
- ADVIA® Chemistry IRON_2 Reagents 10341118Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall0 reports
- ADVIA® Chemistry UN Urea Nitrogen Reagents 10309051Siemens Healthcare Diagnostics Inc. · UREASE AND GLUTAMIC DEHYDROGENASE, UREA NITROGEN1 recall0 reports
- Atellica CH Ca 11097595Siemens Healthcare Diagnostics Inc. · CRESOLPHTHALEIN COMPLEXONE, CALCIUM14 reports
- ADVIA® Chemistry CRE_2c Creatinine Reagents, Concentrated 10283346Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE1 report
- Atellica CH IP 11097611Siemens Healthcare Diagnostics Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page