Skip to content

HANA/PROFx CLASSIC FEMORAL HOOK RIGHT REF 6850-144 recallZ-1573-2022 · Mizuho OSI · initiated Jul 29, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mizuho OSI or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One batch of "regular" femoral hooks were incorrectly etched with the "classic" femoral hook model number.

FDA root cause category: Process control

What the company is doing

On or about 07/05/2022, Muzuho OSI Representative started visiting customers to inform them of a labeling error with the Femur Hook, Right 6850-918 which was incorrectly label with part number 6850-144. Mizuho Representatives were instructed to inform customers: -Over this summer, their facility should have received a Classic Femur Hook, Right 6850-144, either as part of your Hana Table purchase or as a replacement part. But, in error, may have received the other style of femur hook mislabeled with the part number 6850-144. -Hand deliver the Customer Notification Letter and Acknowledge Form. Mizuho Representative are further instructed to: -Inspect the customers' inventory and swap/replace all incorrectly labeled products with the correctly labeled product.

Recall details

Recall number
Z-1573-2022
Status
Open, Classified
Initiated by the company
Jul 29, 2021
Recalling firm
Mizuho OSI
Quantity
13 devices
Product code
LXH
Distribution
U.S Nationwide Distribution .: CA, FL, GA, MI, NE, NJ, TN, UT, and WA O.U.S.: Canada

Product description

HANA/PROFx CLASSIC FEMORAL HOOK RIGHT REF 6850-144

Code information (models, lots, serial numbers)

Model Number: 6850-144 UDI Code: (01)00842430106842(10)MIZ210712(11)210712 Lot/Batch Number: MIZ210712

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page