Skip to content

Metalogix Self-Drill Half-Pin, a component of the… recallZ-1649-2022 · New Standard Device Inc · initiated Dec 28, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from New Standard Device Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Complaints were received for Self-drill Half-Pin tip breakages occurring during surgery.

FDA root cause category: Device Design

What the company is doing

On 12/28/2020, Metalogix send a "Field Notice: Self-Drill Half-Pin" via email. The Notice only informed customer that the Citieffe Half-Pins will temporarily be used to replace the current 5mm Self-Drill Half-Pins. A contact number for Mr. Elton Ely, QA Manager was provided in the notice 833-659-2019 ext. 6

Recall details

Recall number
Z-1649-2022
Status
Open, Classified
Initiated by the company
Dec 28, 2020
Recalling firm
New Standard Device Inc
Quantity
410 total devices
Product code
KTT
FDA submissions
K181630
Distribution
U.S.: AZ, GA, IL MT, NC, NJ, NV, OH, PA, TX, and VA O.U.S.: None

Product description

Metalogix Self-Drill Half-Pin, a component of the Revolution External Fixation System. Orthopedic. Metalogix Self-Drill Half-Pin 5x30x215mm REF 620030 Metalogix Self-Drill Half-Pin 5x40x215mm REF 620040 Metalogix Self-Drill Half-Pin 5x60x215mm REF 620060 Metalogix Self-Drill Half-Pin 5x80x215mm REF 620080

Code information (models, lots, serial numbers)

Part Number / DESCRIPTION / UDI-DI Code / Lot Numbers: 620030 / Self-Drill Half-Pin 5x30x215mm / 854641008198 / A29418X, A17219CA 620040 / Self-Drill Half-Pin 5x40x215mm / 854641008204 / A29418Y, A17219CB 620060 / Self-Drill Half-Pin 5x60x215mm / 854641008211 / A29418Z, A17219CC 620080 / Self-Drill Half-Pin 5x80x215mm / 854641008228 / A29518A, A17219CD

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K181630 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page