VITROS recallZ-1675-2022 · Ortho-Clinical Diagnostics, Inc. · initiated Jul 19, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.

FDA root cause category: Under Investigation by firm

What the company is doing

Ortho CLinical issued Urgent Medical Device Correction letter on 19 July 2022 to consignees who received a VITROS XT3400 or a VITROS XT7600 System notified via (CL2020-176) of the issue and instructed to: " If your VITROS XT 3400 and/or VITROS XT 7600 Systems require a system shutdown, routine Quality Control processing for MicroSlide assays must be performed per your laboratory procedure, prior to processing patient samples, to determine if the anomaly has occurred. " If QC results are outside your established ranges, Ortho recommends performing a second system shutdown and re-processing QC fluids for those assays that were outside your established ranges. If QC results are still outside established ranges following a second system shutdown, contact Ortho Care" Technical Solutions Center for guidance. " Ortho recommends using the same vial of QC fluids (where possible) throughout these steps to determine if the anomaly has occurred. Refer to the control storage and stability documentation for fluid stability guidelines once opened or reconstituted. " Post this notification by each VITROS XT 3400 and/or XT 7600 System until this issue is resolved. Ortho is currently investigating this issue; a follow-up communication will be released once our investigation is complete. " Complete the enclosed Confirmation of Receipt no later than July 26, 2022. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Recall details

Recall number
Z-1675-2022
Status
Open, Classified
Initiated by the company
Jul 19, 2022
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
506 units
Product code
JJE
FDA submissions
K182063
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

Product description

VITROS XT 3400 Chemistry System with Version 3.7.2 and below Product Code: 6844458

Code information (models, lots, serial numbers)

UDI-DI: 10758750031986 Systems start with 3400XXXX where XXXX is unique for each analyzer

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90681).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page