Insertion Kit for use with SENSATION PLUS 8F. 50cc… recallZ-1687-2022 · Datascope Corporation · initiated Aug 5, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain lots containing undersized dilator.
FDA root cause category: Vendor change control
What the company is doing
The firm distributed an Urgent Medical Device Removal notification, by mail, dated August 5, 2022. The notification informs customers that certain lots of 8Fr. 50cc IAB catheter and insertion kits are being recalled due to an undersized vessel dilator. During a sheathless insertion, the undersized inner diameter may result in the inability to insert the guidewire through the inner lumen of the dilator and the undersized outer diameter may reduce dilation of the vessel. If used in a procedure, recalled product could result in a procedural delay, gas emoli, and vessel injury. Customers are asked to examine their inventory to determine if they have nay recalled product and to quarantine any affected units. Customers with recalled product are to contact Datascope Corp./Getinge Customer Service at 888-9GETUSA (888-943-8872, option 2) between the hours of 8 AM and 6 PM to request a return authorization number to receive full credit. Distributors who have shipped any product are to forward the notification to customers. The recall notification is to be forwarded to all users at hospitals/facilities. Customers are asked to complete and return the recall response form by email to [email protected] or by fax to 877-647-1921.
Recall details
- Recall number
- Z-1687-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 5, 2022
- Posted by the FDA
- Sep 8, 2022
- Recalling firm
- Datascope Corporation
- Quantity
- 10,427 units
- Product code
- DSP
- FDA submissions
- K133074
- Distribution
- Awaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Iceland, Ireland, Italy, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Product description
Insertion Kit for use with SENSATION PLUS 8F. 50cc Intra-Aortic Balloon Catheter
Code information (models, lots, serial numbers)
Code information reads "UDI-DI, Lot No., Finished Goods Part No.": 10607567108599, 3000170709, 0884-00-0019-23; 10607567108599, 3000182611, 0884-00-0019-23; 10607567108599, 3000202083, 0884-00-0019-23
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Insertion Kit with 8Fr. 6” (15cm) Introducer for Use with the SENSATION PLUS 8Fr 0884-00-0019-23Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90684).
- Z-1685-2022 · Open, ClassifiedMEGA 8Fr 50cc Intra-Aortic Balloon Catheter
- Z-1686-2022 · Open, ClassifiedSensation Plus 8Fr 50cc Intra-Aortic Balloon Catheter
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page