Skip to content

ARCHITECT recallZ-1660-2022 · Abbott Laboratories · initiated Jul 26, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.

FDA root cause category: Under Investigation by firm

What the company is doing

On July 26, 2022, Abbott issued a Product Recall letter for this issue sent the letter via FedEx priority overnight express. The letter provided the reason for the recall and hazard involved and requested the consignee take the following actions: "" IMMEDIATELY discontinue use of ARCHITECT STAT Troponin-I, List Number 2K41-27 lot number 63718UN22 and destroy all inventory of lot 63718UN22 according to your laboratory procedures. " Please contact Customer Support to order replacement material. " Please review this letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for review of previously reported patient results using lot number 63718UN22. " Complete and return the Customer Reply Form. " If you have forwarded the product listed above to other laboratories, please inform them of this Product Recall and provide to them a copy of this letter. " Please retain this letter for your laboratory records." Questions regarding the information were provided the following contact information: If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week).

Recall details

Recall number
Z-1660-2022
Status
Open, Classified
Initiated by the company
Jul 26, 2022
Recalling firm
Abbott Laboratories
Quantity
2451 kits of 100 tests
Product code
MMI
FDA submissions
K041192
Distribution
Nationwide

Product description

ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27

Code information (models, lots, serial numbers)

UDI (01) 00380740048525 (17) 991231 (10) 12345M100 (240) 2K4127 Lot lot 63718UN22

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page