Surgitube recallZ-1799-2022 · Mckesson Medical-Surgical Inc. Corporate Office · initiated May 25, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mckesson Medical-Surgical Inc. Corporate Office or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.
FDA root cause category: Environmental control
What the company is doing
Mckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 [email protected]. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent possible patient harm. Please complete and return the enclosed response form as soon as possible even if you do not have any product on hand. Customers should contact Sedgwick directly by phone at 1-888-943-2401 to return the recalled products or if they have additional questions regarding this notification. Representatives are available Monday Friday, 8:00 am 5:00 pm PT
Recall details
- Recall number
- Z-1799-2022
- Status
- Open, Classified
- Initiated by the company
- May 25, 2022
- Posted by the FDA
- Sep 26, 2022
- Recalling firm
- Mckesson Medical-Surgical Inc. Corporate Office
- Quantity
- a. 17 b. 7 c. 1 d. 2 e. 5
- Product code
- GER
- Distribution
- US Nationwide distribution.
Product description
a. DERMA SCIENCES, INC Surgitube BANDAGE, GAUZE SURGI WHT SZ15/8" 50YDS Catalog #GL-219. b. DERMA SCIENCES, INC Surgitube BANDAGE, GAUZE SURGI WHT SZ2 1" 50YDS Catalog # GL-220. c. DERMA SCIENCES, INC Surgitube BANDAGE, GAUZE SURGI WHT SZ4 25/8" 50YDS Catalog # GL-222. d. DERMA SCIENCES, INC Surgitube BANDAGE, GAUZE WHT W/O APPL 5/8"X50YDS Catalog # GL-241. e. DERMA SCIENCES, INC Surgitube BANDAGE, GAUZE WHT W/O APPL 11/2"X50YDS Catalog # GL-243
Code information (models, lots, serial numbers)
GTIN: a. 809958001097. b. 809958001103. c. 00809958001127. d. 00809958001141. e. 00809958001165
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Surgitube® 219Derma Sciences Canada Inc · BANDAGE, ELASTIC1 recall0 reports
- Surgitube® 220Derma Sciences Canada Inc · BANDAGE, ELASTIC1 recall0 reports
- Surgitube® 222Derma Sciences Canada Inc · BANDAGE, ELASTIC1 recall0 reports
- Surgitube® 241Derma Sciences Canada Inc · BANDAGE, ELASTIC1 recall0 reports
- Surgitube® 243Derma Sciences Canada Inc · BANDAGE, ELASTIC1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90731).
- Z-1787-2022 · Open, ClassifiedDERMARITE INDUSTRIES SafeWash Saline Saline Wound Flush SafeWash 7.1 oz. Can Sterile 0.9% Sodium Chloride Model Number: 00245
- Z-1788-2022 · Open, Classifieda. Skin Adhesive McKesson LiquiBand Exceed 0.4 mL Liquid Dome Applicator Tip 2-Octyl Cyanoacrylate. Model Number: 122-LBXS. b. Skin Adhesive McKesson LiquiBand Exceed 0.8 mL Liqui…
- Z-1789-2022 · Open, ClassifiedDERMARITE INDUSTRIES DermaCol 100 Collagen Powder DermaCol 100 Collagen 1 Gram Model Number: 00300E
- Z-1790-2022 · Open, Classifieda. SporView Culture Set SporView Sterilization Biological Indicator Kit Steam / EO Gas / Dry Heat / Chemical Vapor Model Number: CS-020 b. SPORVIEW CULTURE SET SporView Steriliza…
- Z-1791-2022 · Open, ClassifiedSPS MEDICAL SUPPLY Crosstex International Sterilization Roll Sani-Roll Ethylene Oxide (EO) Gas / Steam 3 Inch X 100 Foot Transparent / Blue Self Seal Paper / Film Model Number: SCT…
- Z-1792-2022 · Open, Classifieda. SPS MEDICAL SUPPLY Steam Indicator Strip SPSmedical Sterilization Chemical Indicator Strip Steam 4 Inch Model Number: SIS-250. b. SPS MEDICAL SUPPLY SteamPlus Sterilization In…
- Z-1793-2022 · Open, Classifieda. STERIS CORP Celerity Celerity 20 Sterilization Indicator Challenge Pack Steam 3 Inch Length Model Number: LCB053. b. STERIS CORP VERIFY Verify SixCess Sterilization Flash Indica…
- Z-1794-2022 · Open, Classifieda. Comply INDICATOR, COMPLY VAPOR GAS (250/BG 4BG/CS) 3M Model Number: 1248. b. 3M Attest INDICATOR, CHEMICAL VAPORIZEDHYDROGEN PEROXIDE (250/BG) Model Number: 1348.
- Z-1795-2022 · Open, ClassifiedCooper Surgical PREFILTER, SMOKE EVAC (10/BX)CPRSRG Model Number: 52560.
- Z-1796-2022 · Open, Classifieda. DERMARITE INDUSTRIES DermaCol Collagen Dressing DermaCol Without Border Collagen / Sodium Alginate / Carboxyl Methylcellulose (CMC) / Ethylenediamine-tetraacetic Acid (EDTA) 4 X…
- Z-1797-2022 · Open, ClassifiedMERIT MEDICAL SYSTEMS StayFIX FIXATION DEVICE, STAYFIX F/PERCUTAN CATH 5FR-14FR Model Number: 884450023358. Used to secure anaesthesia catheter.
- Z-1800-2022 · Open, Classifieda. INTEGRA LIFESCIENCE (WND CARE) Shur-Conform Oil Emulsion Oil Emulsion Impregnated Dressing Shur-Conform 3 X 8 Inch Knitted Cellulose Acetate Petrolatum Emulsion Sterile Model N…
- Z-1801-2022 · Open, ClassifiedSkin markers: a. IZI MEDICAL PRODUCTS LLC Multi-Modality Markers SKIN MARKER RADIO 1.5MM 50BX Catalog #MM3005. b. SKIN MARKER MAMMO BB2.0 100B - Mammography Skin Marker Catalog…
- Z-1803-2022 · Open, ClassifiedAbsorbent dressings. a. DERMA SCIENCES, INC. McKesson PACKING STRIP, PLAIN 1/4"X5YDSSTR LF (12 MGM61 61- Catalog #61-59120. b. DERMA SCIENCES, INC. McKesson PACKING STRIP, PLAIN 1/…
- Z-1805-2022 · Open, Classifieda. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, CALCIUM ALGINATE SHEET 4"X8" (5/BX 10BX/CS), Catalog #3563. b. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, CALCIUM ALGINATE AG2X…
- Z-1806-2022 · Open, Classifieda. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WOUND GELLING FIBERHIGH-ABSRB 4"X4 3/4" (10/BX) Catalog #87400. b. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WND GELLING FI…
- Z-1807-2022 · Open, Classified3M SYSTAGENIX/KCI KerraCel Ag DRESSING, WOUND KERRACEL AG 4X5 (10/CT 10CT/CS) Catalog #CWL1158
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page