MEDLINE CLEARPRO TRACH DOUBLE SWIVEL CLOSED SUCTION… recallZ-1856-2022 · MEDLINE INDUSTRIES, LP - Northfield · initiated Aug 3, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
FDA root cause category: Under Investigation by firm
What the company is doing
On about 08/03/2022, Medline notified consignees via first class mail letter of the recall. The letter instructed consignees to immediately review inventory and destroy all affected product and complete the response form indicated at the link provided in the letter. Distributors were instructed to notify customers if further distributed, instruct those customers destroy any affected product, and include customer quantities on response form.
Recall details
- Recall number
- Z-1856-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 3, 2022
- Posted by the FDA
- Sep 29, 2022
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 34,920 units
- Product code
- BSY
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Panama, Qatar, United Arab Emirates
Product description
MEDLINE CLEARPRO TRACH DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12T b. TRACH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14T
Code information (models, lots, serial numbers)
a. UDI-DI (GTIN) case: (01)20888277703374; UDI-DI (GTIN) Each: (01)10888277703377, Lot Numbers 6921090011, 6922030011; b. UDI-DI (GTIN) case: (01)20888277703398; UDI-DI (GTIN) Each: (01)10888277703391, Lot Numbers 6921040012, 6921070011, 6921080011, 6921100011, 6921120011, 6921120021, 6922030011, 6922040011
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline DYNCPDS14TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall3 reports
- Medline DYNCPDS12TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90722).
- Z-1855-2022 · Open, ClassifiedMEDLINE CLEARPRO DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. ET LENGTH DO…
- Z-1857-2022 · Open, ClassifiedMEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIEC…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page