Surpria: Software Version V3.11, V3.22 recallZ-1749-2022 · FUJIFILM Healthcare Americas Corporation · initiated Jul 12, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from FUJIFILM Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation
FDA root cause category: Under Investigation by firm
What the company is doing
FujiFilm issued Urgent Medical Device Letter were sent via email on July 12, 2022, consignees without an email contact were notified via FedEx on July 12, 2022 and July 15, 2022. Letter states reason for recall, health risk and action to take: Until your upgrade is scheduled and completed, you can continue to safely use your CT system by following these additional instructions: 1. When entering alphabetic characters in the patient information, please enter them in capital letters (e.g. ready READY ) 2. Set Auto Capital Letter to ON in the Preference Settings window. This will convert all text to capital letters for certain patient information fields. 3. Please register the following information in capital letters, as Auto Capital Letter does not apply to the following fields: a. (0008,0050) Accession Number b. (0008,0090) Referring Physician s Name c. (0010,0010) Patient s Name d. (0010,0020) Patient ID e. (0010,4000) Patient Comments f. (0032,1032) Requesting Physician g. (0032,1033) Requesting Service h. (0032,1060) Requested Procedure Description i. (0032,1070) Requested Contrast Agent j. (0040,0006) Scheduled Performing Physician s Name Please complete and return the Field Action Verification Form attached on page 5 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Fujifilm regrets any inconvenience this action may cause and appreciates your understanding and cooperation. Please contact our field action coordinator, Jeffrey Wan, via phone at 201-675-8947 or email at [email protected]
Recall details
- Recall number
- Z-1749-2022
- Status
- Open, Classified
- Initiated by the company
- Jul 12, 2022
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Quantity
- 20 units
- Product code
- JAK
- FDA submissions
- K183291
- Distribution
- US Nationwide Distribution
Product description
Surpria: Software Version V3.11, V3.22
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI: (01)04580292766601(11)200130(21)W5067(250)G560045319 (01)04580292766601(11)200220(21)W5068(250)G560048119 (01)04580292766601(11)200902(21)W5069(250)G560071519 (01)04580292766601(11)201210(21)W5070(250)G560075519 (01)04580292766601(11)201225(21)W5071(250)G1Z0040620 (01)04580292766601(11)210218(21)W5072(250)G1Z0063520 (01)04580292766601(11)210512(21)W5073(250)G1Z0244920 (01)04580292766601(11)210519(21)W5074(250)G1Z0048121 (01)04573596200520(11)210818(21)W5075(250)G1Z0083221 (01)04573596200520(11)210903(21)W5076(250)G1Z0094621 (01)04573596200520(11)210903(21)W5077(250)G1Z0094521 (01)04573596200520(11)210903(21)W5077(250)G1Z0094521 (01)04573596200520(11)211018(21)W5079(250)G1Z0123621 (01)04573596200520(11)211115(21)W5080(250)G1Z0190221 (01)04573596200520(11)211122(21)W5081(250)G1Z0190021 (01)04573596200520(11)211122(21)W5082(250)G1Z0190121 (01)04573596200520(11)211227(21)W5083(250)G1Z0190321 (01)04573596200520(11)211228(21)W5084(250)G1Z0210821 (01)04573596200520(11)220209(21)W5085(250)G1Z0210921 (01)04573596200520(11)220215(21)W5086(250)G1Z0211021 Serial Numbers: G560045319 G560048119 G560071519 G560075519 G1Z0040620 G1Z0063520 G1Z0244920 G1Z0048121 G1Z0083221 G1Z0094621 G1Z0094521 G1Z0115321 G1Z0123621 G1Z0190221 G1Z0190021 G1Z0190121 G1Z0190321 G1Z0210821 G1Z0210921 G1Z0211021
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Hitachi SupriaHitachi, Ltd. · System, X-Ray, Tomography, Computed2 recalls2 reports
- SupriaFujifilm Corporation · System, X-Ray, Tomography, Computed2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90751).
- Z-1748-2022 · Open, ClassifiedSCENARIA View: Software Version V1.08, V1.09B, V1.09C, V1.09D, V1.0B, V1.0C
- Z-1750-2022 · Open, ClassifiedSurpria 64: Software Version V3.11, V3.22
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page