Skip to content

Grappler Suture Anchor System recallZ-1776-2022 · Paragon 28, Inc. · initiated Jul 13, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Paragon 28, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Suture anchor may break upon insertion, which could lead to potential debris in patient, and delays in surgery.

FDA root cause category: Device Design

What the company is doing

On 7/13/22, recall notices were emailed to distributors who were asked to do the following: 1) Review physical inventory of product within your control and at a hospital/health care facility within your region. Any affected product that is identified is to be discontinued, quarantined and not distributed any further. 2) Return affected devices to the recalling firm. 3) Complete and return the Recall Response form. For inquiries, contact the recalling firm at [email protected]

Recall details

Recall number
Z-1776-2022
Status
Open, Classified
Initiated by the company
Jul 13, 2022
Posted by the FDA
Sep 22, 2022
Recalling firm
Paragon 28, Inc.
Quantity
145
Product code
MBI
FDA submissions
K211002
Distribution
US Nationwide distribution in the states of AZ, IN, LA, NJ, TX, WA, NY, MD, WI, OK, KY, MI, OH, CA, FL, KS, PA.

Product description

Grappler Suture Anchor, Titanium, 3.0 x 10mm, Sterile Kit, REF: P44-110-3010-SK, for use within the Grappler Suture Anchor System

Code information (models, lots, serial numbers)

UDI-DI: 00889795114228, Lot: 5007499

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates